Index: NVDataMgmt.c =================================================================== diff -u -r2b2282f2c9eae29bf860b3154cabd0289115551b -r15388d4947d9ac8dbf0a7e9356e953fef126fbb2 --- NVDataMgmt.c (.../NVDataMgmt.c) (revision 2b2282f2c9eae29bf860b3154cabd0289115551b) +++ NVDataMgmt.c (.../NVDataMgmt.c) (revision 15388d4947d9ac8dbf0a7e9356e953fef126fbb2) @@ -133,6 +133,7 @@ #define DEFAULT_ENABLE_CHEM_DISINFECT 0 ///< Disable chemical disinfect in institutional record. #define DEFAULT_MAX_RO_REJECTION_RATIO_PCT 90 ///< Max RO rejection ratio in percent. #define DEFAULT_MIN_INLET_WATER_COND_ALARM_US_P_CM 200.0F ///< Min inlet water conductivity alarm limit in uS/cm. +#define DEFAULT_MIN_VEN_WIDE_WINDOW_LIMIT_MMHG 5 ///< Min venous wide window limit in mmHG. #endif // DG specific defines @@ -3973,6 +3974,7 @@ hdInstitutionalGroup.hdInstitutionalRecord.enableChemicalDisinfect = DEFAULT_ENABLE_CHEM_DISINFECT; hdInstitutionalGroup.hdInstitutionalRecord.minRORejectionRatioPCT = DEFAULT_MAX_RO_REJECTION_RATIO_PCT; hdInstitutionalGroup.hdInstitutionalRecord.minInletWaterCondAlarmLimitUSPCM = DEFAULT_MIN_INLET_WATER_COND_ALARM_US_P_CM; + hdInstitutionalGroup.hdInstitutionalRecord.minVenPressWideLimitWindowMMHG = DEFAULT_MIN_VEN_WIDE_WINDOW_LIMIT_MMHG; hdInstitutionalGroup.hdInstitutionalRecord.calibrationTime = getRTCTimestamp(); hdInstitutionalGroup.hdInstitutionalRecord.crc = crc16 ( (U08*)&hdInstitutionalGroup.hdInstitutionalRecord, sizeof( HD_INSTITUTIONAL_RECORD_T ) - sizeof( U16 ) );