Index: firmware/App/Controllers/BloodFlow.c =================================================================== diff -u -rfa0ab419491364c1d2a28caf216e20175181372d -r6fcdec8a308be986e5a64b4222918e718622a19d --- firmware/App/Controllers/BloodFlow.c (.../BloodFlow.c) (revision fa0ab419491364c1d2a28caf216e20175181372d) +++ firmware/App/Controllers/BloodFlow.c (.../BloodFlow.c) (revision 6fcdec8a308be986e5a64b4222918e718622a19d) @@ -46,8 +46,8 @@ #define MAX_BLOOD_PUMP_PWM_STEP_UP_CHANGE 0.008 ///< Max duty cycle change when ramping up ~ 100 mL/min/s. #define MAX_BLOOD_PUMP_PWM_STEP_DN_CHANGE 0.016 ///< Max duty cycle change when ramping down ~ 200 mL/min/s. -#define MAX_BLOOD_PUMP_PWM_DUTY_CYCLE 0.88 ///< Controller will error if PWM duty cycle > 90%, so set max to 88% -#define MIN_BLOOD_PUMP_PWM_DUTY_CYCLE 0.12 ///< Controller will error if PWM duty cycle < 10%, so set min to 12% +#define MAX_BLOOD_PUMP_PWM_DUTY_CYCLE 0.89 ///< Controller will error if PWM duty cycle > 90%, so set max to 89% +#define MIN_BLOOD_PUMP_PWM_DUTY_CYCLE 0.10 ///< Controller will error if PWM duty cycle < 10%, so set min to 10% /// Interval (ms/task time) at which the blood pump is controlled. static const U32 BP_CONTROL_INTERVAL = ( 10000 / TASK_GENERAL_INTERVAL ); @@ -82,14 +82,19 @@ #define BP_MAX_CURR_WHEN_STOPPED_MA 150.0 ///< Motor controller current should not exceed this when pump should be stopped #define BP_MAX_CURR_WHEN_RUNNING_MA 2000.0 ///< Motor controller current should not exceed this when pump should be running #define BP_MAX_CURR_ERROR_DURATION_MS 2000 ///< Motor controller current errors persisting beyond this duration will trigger an alarm - -#define BP_SPEED_ADC_TO_RPM_FACTOR 1.280938 ///< Conversion factor from ADC counts to RPM for blood pump motor + +#ifndef V2_0_SYSTEM + #define BP_SPEED_ADC_TO_RPM_FACTOR 1.751752 ///< Conversion factor from ADC counts to RPM for blood pump motor + #define BP_MOTOR_RPM_TO_PWM_DC_FACTOR 0.00025 ///< ~40 BP motor RPM = 1% PWM duty cycle +#else + #define BP_SPEED_ADC_TO_RPM_FACTOR 1.280938 ///< Conversion factor from ADC counts to RPM for blood pump motor + #define BP_MOTOR_RPM_TO_PWM_DC_FACTOR 0.0003125 ///< ~32 BP motor RPM = 1% PWM duty cycle +#endif #define BP_CURRENT_ADC_TO_MA_FACTOR 3.002 ///< Conversion factor from ADC counts to mA for blood pump motor #define BP_REV_PER_LITER 150.0 ///< Rotor revolutions per liter #define BP_ML_PER_MIN_TO_PUMP_RPM_FACTOR ( BP_REV_PER_LITER / ML_PER_LITER ) ///< Conversion factor from mL/min to motor RPM. #define BP_GEAR_RATIO 32.0 ///< Blood pump motor to blood pump gear ratio -#define BP_MOTOR_RPM_TO_PWM_DC_FACTOR 0.00028 ///< ~28 BP motor RPM = 1% PWM duty cycle #define BP_PWM_ZERO_OFFSET 0.1 ///< 10% PWM duty cycle = zero speed /// Conversion factor from mL/min to estimated PWM duty cycle %. #define BP_PWM_FROM_ML_PER_MIN(rate) ( (rate) * BP_ML_PER_MIN_TO_PUMP_RPM_FACTOR * BP_GEAR_RATIO * BP_MOTOR_RPM_TO_PWM_DC_FACTOR + BP_PWM_ZERO_OFFSET )