Index: firmware/App/Controllers/BloodLeak.c =================================================================== diff -u -r4cfe2a27c1c066fa8bda9689c28aec895d7de84b -r076991b7551d319aa510b09247c4c5f77444808f --- firmware/App/Controllers/BloodLeak.c (.../BloodLeak.c) (revision 4cfe2a27c1c066fa8bda9689c28aec895d7de84b) +++ firmware/App/Controllers/BloodLeak.c (.../BloodLeak.c) (revision 076991b7551d319aa510b09247c4c5f77444808f) @@ -337,7 +337,9 @@ } else { +#ifndef IGNORE_BLOOD_LEAK_ALARM activateAlarmNoData( ALARM_ID_HD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE ); +#endif } } else @@ -481,7 +483,7 @@ { BLOOD_LEAK_STATES_T state = BLOOD_LEAK_SELF_TEST_STATE; -#ifndef IGNORE_BLOOD_LEAK_ALARM +#ifndef IGNORE_BLOOD_LEAK_SELF_TEST if ( SELF_TEST_STATUS_IN_PROGRESS == bloodLeakSelfTestStatus ) { if ( FALSE == noFPGABloodLeakDetected() ) // Faked blood leak caused by independent MCU board @@ -492,8 +494,9 @@ else if ( TRUE == didTimeout( bloodLeakSelfTestStartTime, BLOOD_LEAK_TIMEOUT_MS ) ) { bloodLeakSelfTestStatus = SELF_TEST_STATUS_FAILED; - +#ifndef IGNORE_BLOOD_LEAK_ALARM activateAlarmNoData( ALARM_ID_HD_BLOOD_LEAK_SELF_TEST_FAILURE ); +#endif } } else