Index: firmware/App/Controllers/DGInterface.c =================================================================== diff -u -rd23f94b810d65dd54809b0bfeb32f0f763051d4d -r7a1c1fe89abb13803db1961146b7b0d97574de16 --- firmware/App/Controllers/DGInterface.c (.../DGInterface.c) (revision d23f94b810d65dd54809b0bfeb32f0f763051d4d) +++ firmware/App/Controllers/DGInterface.c (.../DGInterface.c) (revision 7a1c1fe89abb13803db1961146b7b0d97574de16) @@ -49,7 +49,6 @@ #define DIALYSATE_TEMP_LOW_SAFETY_LIMIT_C 42.0F ///< Dialysate low safety temperature limit in C. #define DIALYSATE_TEMP_LOW_SAFETY_TIMEOUT_MS ( 10 * MS_PER_SECOND ) ///< Dialysate temperature low safety timeout in milliseconds. #define DG_DATA_FRESHNESS_TIMEOUT_MS ( 3 * MS_PER_SECOND ) ///< DG data freshness timeout (in ms). -#define DIALYSATE_FLOW_FRESHNESS_TIMEOUT_MS ( 60 * MS_PER_SECOND ) ///< Dialysate flow freshness timeout (in ms). // DN-31OCT2022 // ********** private data ********** @@ -191,7 +190,6 @@ initPersistentAlarm( ALARM_ID_HD_NEW_DIALYSATE_TEMPERATURE_DATA_MESSAGE_NOT_RECEIVE, DG_DATA_FRESHNESS_TIMEOUT_MS, DG_DATA_FRESHNESS_TIMEOUT_MS ); initPersistentAlarm( ALARM_ID_HD_NEW_RESERVOIRS_DATA_MESSAGE_NOT_RECEIVE, DG_DATA_FRESHNESS_TIMEOUT_MS, DG_DATA_FRESHNESS_TIMEOUT_MS ); initPersistentAlarm( ALARM_ID_HD_NEW_DG_OPERATION_MODE_MESSAGE_NOT_RECEIVE, DG_DATA_FRESHNESS_TIMEOUT_MS, DG_DATA_FRESHNESS_TIMEOUT_MS ); - initPersistentAlarm( ALARM_ID_HD_DIAL_IN_FLOW_CHECK_FAILURE, DIALYSATE_FLOW_FRESHNESS_TIMEOUT_MS, DIALYSATE_FLOW_FRESHNESS_TIMEOUT_MS ); // DN-31OCT2022 } /**********************************************************************//** @@ -248,11 +246,14 @@ // Trigger alarm if not receiving new DG op mode message in timely manner checkDGDataFreshness( ALARM_ID_HD_NEW_DG_OPERATION_MODE_MESSAGE_NOT_RECEIVE, &dgOpModeDataFreshFlag ); +<<<<<<< HEAD // Check to see if DG has restarted*/ // Trigger alarm if not receiving new DG op mode message in timely manner //checkDGDataFreshness( ALARM_ID_HD_DIAL_IN_FLOW_CHECK_FAILURE, &dgDialysateFlowDataFreshFlag ); // DN-31OCT2022 - TODO: Will remove this line??? +======= +>>>>>>> DEN-14150-hd-dialysate-flow-estimation // Check to see if DG has restarted checkDGRestart();