Index: firmware/App/Controllers/DGInterface.c =================================================================== diff -u -re14e14eedeb52b901618222bb59320b955bf7f5a -r710549dd068ceab73277d24ce4a546f99c0ded8d --- firmware/App/Controllers/DGInterface.c (.../DGInterface.c) (revision e14e14eedeb52b901618222bb59320b955bf7f5a) +++ firmware/App/Controllers/DGInterface.c (.../DGInterface.c) (revision 710549dd068ceab73277d24ce4a546f99c0ded8d) @@ -961,8 +961,8 @@ * @brief * The checkDialysateTemperature function checks the dialysate temperature * reported by DG and alarm if temperature is out of range. - * @details Inputs: dgTrimmerTempSet, dgDialysateTemp, dgRedundantDialysateTemp - * @details Outputs: alarm if dialysate temperature is out of accepted range + * @details Inputs: dgTrimmerTempSet, dgDialysateTemp + * @details Outputs: none * @return none *************************************************************************/ void checkDialysateTemperature( void ) @@ -979,7 +979,7 @@ #endif { checkPersistentAlarm( ALARM_ID_HD_DIALYSATE_TEMP_OUT_OF_HIGH_SAFETY_RANGE, isTDiTempAboveHighSafety, dgDialysateTemp, DIALYSATE_TEMP_HIGH_SAFETY_LIMIT_C ); - checkPersistentAlarm( ALARM_ID_HD_DIALYSATE_TEMP_OUT_OF_LOW_SAFETY_RANGE, isTDiTempAboveLowSafety, dgDialysateTemp, DIALYSATE_TEMP_LOW_SAFETY_LIMIT_C ); + checkPersistentAlarm( ALARM_ID_HD_DIALYSATE_TEMP_OUT_OF_LOW_SAFETY_RANGE, isTDiTempAboveLowSafety, dgDialysateTemp, DIALYSATE_TEMP_LOW_SAFETY_LIMIT_C ); checkPersistentAlarm( ALARM_ID_HD_DIALYSATE_TEMP_ABOVE_TARGET_TEMP, isTDiTempAboveDialysateTarget, dgDialysateTemp, TDiHigh ); checkPersistentAlarm( ALARM_ID_HD_DIALYSATE_TEMP_BELOW_TARGET_TEMP, isTDiTempBelowDialysateTarget, dgDialysateTemp, TDiLow );