Index: firmware/App/Controllers/DialInFlow.c =================================================================== diff -u -r09e6cf9de34acf18f6e1138bf56ac0edb4821186 -rf7cdfd3256f195cd8aa06e28009cd5f26b50f08c --- firmware/App/Controllers/DialInFlow.c (.../DialInFlow.c) (revision 09e6cf9de34acf18f6e1138bf56ac0edb4821186) +++ firmware/App/Controllers/DialInFlow.c (.../DialInFlow.c) (revision f7cdfd3256f195cd8aa06e28009cd5f26b50f08c) @@ -100,7 +100,7 @@ #define DIP_GEAR_RATIO 32.0 ///< DialIn pump motor to dialIn pump gear ratio. #define DIP_PWM_ZERO_OFFSET 0.1 ///< 10% PWM duty cycle = zero speed. /// Macro converts flow rate to estimate PWM needed to achieve it. // TODO - I added 1.2 gain based on empirical data -#define DIP_PWM_FROM_ML_PER_MIN(rate) ( (rate) * DIP_ML_PER_MIN_TO_PUMP_RPM_FACTOR * DIP_GEAR_RATIO * DIP_MOTOR_RPM_TO_PWM_DC_FACTOR * 1.2 + DIP_PWM_ZERO_OFFSET ) +#define DIP_PWM_FROM_ML_PER_MIN(rate) ( (rate) * DIP_ML_PER_MIN_TO_PUMP_RPM_FACTOR * DIP_GEAR_RATIO * DIP_MOTOR_RPM_TO_PWM_DC_FACTOR * 1.258 + DIP_PWM_ZERO_OFFSET ) /// Conversion from PWM duty cycle % to commanded pump motor speed. #define DIP_PWM_TO_MOTOR_SPEED_RPM(pwm) ( ((pwm) - DIP_PWM_ZERO_OFFSET) * 4000.0 ) @@ -975,7 +975,9 @@ if ( lastDialInPumpDirectionCount != dirErrorCnt ) { lastDialInPumpDirectionCount = dirErrorCnt; +#ifndef DISABLE_PUMP_DIRECTION_CHECKS SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_PUMP_DIRECTION_STATUS_ERROR, (U32)HD_PUMP_DIALYSATE_INLET_PUMP ) +#endif } #endif