Index: firmware/App/Controllers/DialInFlow.c =================================================================== diff -u -r0d8381c50f7189c4cfbc2794a4aedf297eea7d15 -r7218f3c5a0afef9ebbf47655a00f37fc2eefd0f9 --- firmware/App/Controllers/DialInFlow.c (.../DialInFlow.c) (revision 0d8381c50f7189c4cfbc2794a4aedf297eea7d15) +++ firmware/App/Controllers/DialInFlow.c (.../DialInFlow.c) (revision 7218f3c5a0afef9ebbf47655a00f37fc2eefd0f9) @@ -1323,9 +1323,7 @@ } if ( ctrlMode < NUM_OF_PUMP_CONTROL_MODES ) { - targetDialInFlowRate.ovInitData = targetDialInFlowRate.data; // backup current target flow rate - targetDialInFlowRate.ovData = value; - targetDialInFlowRate.override = OVERRIDE_KEY; + targetDialInFlowRate.data = value; result = setDialInPumpTargetFlowRate( abs(value), dir, (PUMP_CONTROL_MODE_T)ctrlMode ); } } @@ -1335,31 +1333,6 @@ /*********************************************************************//** * @brief - * The testResetTargetDialInFlowRateOverride function resets the override of the - * target dialysate inlet flow rate. - * @details - * Inputs : none - * Outputs : targetDialInFlowRate - * @return TRUE if override reset successful, FALSE if not - *************************************************************************/ -BOOL testResetTargetDialInFlowRateOverride( void ) -{ - BOOL result = FALSE; - - if ( TRUE == isTestingActivated() ) - { - targetDialInFlowRate.data = targetDialInFlowRate.ovInitData; // restore pre-override target flow rate - targetDialInFlowRate.override = OVERRIDE_RESET; - targetDialInFlowRate.ovInitData = 0; - targetDialInFlowRate.ovData = 0; - result = setDialInPumpTargetFlowRate( targetDialInFlowRate.data, dialInPumpDirection, dialInPumpControlMode ); - } - - return result; -} - -/*********************************************************************//** - * @brief * The testResetMeasuredDialInFlowRateOverride function overrides the measured * dialIn flow rate. * @details