Index: firmware/App/Controllers/DialInFlow.c =================================================================== diff -u -r1c628bfd5d6414b74b8cbd083f66839888a8236b -r1ebbe403d260d9f65695c13d6dbafbe1a93cfdde --- firmware/App/Controllers/DialInFlow.c (.../DialInFlow.c) (revision 1c628bfd5d6414b74b8cbd083f66839888a8236b) +++ firmware/App/Controllers/DialInFlow.c (.../DialInFlow.c) (revision 1ebbe403d260d9f65695c13d6dbafbe1a93cfdde) @@ -1089,7 +1089,9 @@ if ( lastDialInPumpDirectionCount != dirErrorCnt ) { lastDialInPumpDirectionCount = dirErrorCnt; +#ifndef DISABLE_PUMP_DIRECTION_CHECKS SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_PUMP_DIRECTION_STATUS_ERROR, (U32)HD_PUMP_DIALYSATE_INLET_PUMP ) +#endif } dipMCDir = ( getMeasuredDialInPumpMCSpeed() >= 0.0 ? MOTOR_DIR_FORWARD : MOTOR_DIR_REVERSE );