Index: firmware/App/Controllers/DialInFlow.c =================================================================== diff -u -r736cc5b56cc9c784ab1d8fc8687a73d190c35759 -r20207f527bfa5ccc19245a61bbe856d3aa4cf8c0 --- firmware/App/Controllers/DialInFlow.c (.../DialInFlow.c) (revision 736cc5b56cc9c784ab1d8fc8687a73d190c35759) +++ firmware/App/Controllers/DialInFlow.c (.../DialInFlow.c) (revision 20207f527bfa5ccc19245a61bbe856d3aa4cf8c0) @@ -51,7 +51,7 @@ #define MAX_DIAL_IN_PUMP_PWM_STEP_UP_CHANGE 0.01064F ///< Max duty cycle change when ramping up. #define MAX_DIAL_IN_PUMP_PWM_STEP_DN_CHANGE 0.016F ///< Max duty cycle change when ramping down. -#define MAX_DIAL_IN_PUMP_PWM_DUTY_CYCLE 0.90F ///< Controller will error if PWM duty cycle > 90%, so set max to 90%. +#define MAX_DIAL_IN_PUMP_PWM_DUTY_CYCLE 0.78F ///< Max duty cycle is 78% to cap pump speed to 3,400 RPM to prevent rotor speed errors. #define MIN_DIAL_IN_PUMP_PWM_DUTY_CYCLE 0.10F ///< Controller will error if PWM duty cycle < 10%, so set min to 10%. #define MIN_DG_DIAL_FLOW_RATE 10.0F ///< Minimum DG Dialysate Flow Rate (mL/Min)