Index: firmware/App/Controllers/DialInFlow.c =================================================================== diff -u -rccfd15568f1e3d304320c2babb2fd4bcf0413304 -re250309026d33ff57fd00d74ccc05b1595610656 --- firmware/App/Controllers/DialInFlow.c (.../DialInFlow.c) (revision ccfd15568f1e3d304320c2babb2fd4bcf0413304) +++ firmware/App/Controllers/DialInFlow.c (.../DialInFlow.c) (revision e250309026d33ff57fd00d74ccc05b1595610656) @@ -257,9 +257,6 @@ // Initialize persistent alarm for flow sensor signal strength too low initPersistentAlarm( ALARM_ID_DIALYSATE_FLOW_SIGNAL_STRENGTH_TOO_LOW, FLOW_SIG_STRGTH_ALARM_PERSIST, FLOW_SIG_STRGTH_ALARM_PERSIST ); - initPersistentAlarm( ALARM_ID_HD_DP_FLOW_READ_TIMEOUT_ERROR, 0, DIALYSATE_FLOW_FAST_READ_TO_PERSIST ); - initPersistentAlarm( ALARM_ID_HD_DP_FLOW_SLOW_READ_TIMEOUT_ERROR, 0, DIALYSATE_FLOW_SLOW_READ_TO_PERSIST ); - initPersistentAlarm( ALARM_ID_HD_DP_FLOW_SENSOR_ERROR, 0, DIALYSATE_FLOW_COMM_ERROR_PERSIST ); initPersistentAlarm( ALARM_ID_HD_DIAL_IN_FLOW_OUT_OF_RANGE, 0, DIP_MAX_FLOW_RATE_OUT_OF_RANGE_PERSIST ); } @@ -443,29 +440,13 @@ dialysateFlowCalRecord.hdFlowSensors[ CAL_DATA_HD_DIALYSATE_FLOW_SENSOR ].offset; #ifndef DISABLE_PUMP_FLOW_CHECKS - if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_HD_DP_FLOW_SENSOR_ERROR, ( flowErrorCtr != lastDialysateFlowCommErrorCount ) ) ) - { - activateAlarmNoData( ALARM_ID_HD_DP_FLOW_SENSOR_ERROR ); - } if ( flowStatus != DFM_SENSOR_CONNECTED_STATUS ) { SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_DIALYSATE_FLOW_STATUS_SELF_TEST_FAILURE, (U32)flowStatus ); } lastDialysateFlowCommErrorCount = flowErrorCtr; #endif -#ifndef DISABLE_FPGA_COUNTER_CHECKS - // Check for stale flow reading - if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_HD_DP_FLOW_READ_TIMEOUT_ERROR, ( fpReadCtr == lastDialysateFlowFastPacketReadCtr ) ) ) - { - activateAlarmNoData( ALARM_ID_HD_DP_FLOW_READ_TIMEOUT_ERROR ); - } - if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_HD_DP_FLOW_SLOW_READ_TIMEOUT_ERROR, ( spReadCtr == lastDialysateFlowSlowPacketReadCtr ) ) ) - { - activateAlarmNoData( ALARM_ID_HD_DP_FLOW_SLOW_READ_TIMEOUT_ERROR ); - } -#endif - // Record flow read counters for next time around lastDialysateFlowFastPacketReadCtr = fpReadCtr; lastDialysateFlowSlowPacketReadCtr = spReadCtr;