Index: firmware/App/Controllers/DialInFlow.c =================================================================== diff -u -re1b010909170d0656a7b87400f295387423fd383 -r4ddbff03c9143b0ece158d927152050d4ae26539 --- firmware/App/Controllers/DialInFlow.c (.../DialInFlow.c) (revision e1b010909170d0656a7b87400f295387423fd383) +++ firmware/App/Controllers/DialInFlow.c (.../DialInFlow.c) (revision 4ddbff03c9143b0ece158d927152050d4ae26539) @@ -5,13 +5,13 @@ * THIS CODE MAY NOT BE COPIED OR REPRODUCED IN ANY FORM, IN PART OR IN * WHOLE, WITHOUT THE EXPLICIT PERMISSION OF THE COPYRIGHT OWNER. * -* @file DialInFlow.c +* @file DialInFlow.c * -* @author (last) H. Nguyen -* @date (last) 28-Oct-2021 +* @author (last) H. Nguyen +* @date (last) 28-Oct-2021 * -* @author (original) Sean -* @date (original) 16-Dec-2019 +* @author (original) Sean +* @date (original) 16-Dec-2019 * ***************************************************************************/ @@ -97,7 +97,7 @@ #define DIP_GEAR_RATIO 32.0 ///< DialIn pump motor to dialIn pump gear ratio. #define DIP_PWM_ZERO_OFFSET 0.1 ///< 10% PWM duty cycle = zero speed. /// Macro converts flow rate to estimate PWM needed to achieve it. // TODO - I added 1.2 gain based on empirical data -#define DIP_PWM_FROM_ML_PER_MIN(rate) ( (rate) * DIP_ML_PER_MIN_TO_PUMP_RPM_FACTOR * DIP_GEAR_RATIO * DIP_MOTOR_RPM_TO_PWM_DC_FACTOR * 1.2 + DIP_PWM_ZERO_OFFSET ) +#define DIP_PWM_FROM_ML_PER_MIN(rate) ( (rate) * DIP_ML_PER_MIN_TO_PUMP_RPM_FACTOR * DIP_GEAR_RATIO * DIP_MOTOR_RPM_TO_PWM_DC_FACTOR * 1.258 + DIP_PWM_ZERO_OFFSET ) /// Conversion from PWM duty cycle % to commanded pump motor speed. #define DIP_PWM_TO_MOTOR_SPEED_RPM(pwm) ( ((pwm) - DIP_PWM_ZERO_OFFSET) * 4000.0 ) @@ -1121,7 +1121,9 @@ if ( lastDialInPumpDirectionCount != dirErrorCnt ) { lastDialInPumpDirectionCount = dirErrorCnt; +#ifndef DISABLE_PUMP_DIRECTION_CHECKS SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_PUMP_DIRECTION_STATUS_ERROR, (U32)HD_PUMP_DIALYSATE_INLET_PUMP ) +#endif } #endif