Index: firmware/App/Controllers/DialInFlow.c =================================================================== diff -u -rfa0ab419491364c1d2a28caf216e20175181372d -r6fcdec8a308be986e5a64b4222918e718622a19d --- firmware/App/Controllers/DialInFlow.c (.../DialInFlow.c) (revision fa0ab419491364c1d2a28caf216e20175181372d) +++ firmware/App/Controllers/DialInFlow.c (.../DialInFlow.c) (revision 6fcdec8a308be986e5a64b4222918e718622a19d) @@ -46,8 +46,8 @@ #define MAX_DIAL_IN_PUMP_PWM_STEP_UP_CHANGE 0.0133 ///< Max duty cycle change when ramping up ~ 200 mL/min/s. #define MAX_DIAL_IN_PUMP_PWM_STEP_DN_CHANGE 0.02 ///< Max duty cycle change when ramping down ~ 300 mL/min/s. -#define MAX_DIAL_IN_PUMP_PWM_DUTY_CYCLE 0.88 ///< Controller will error if PWM duty cycle > 90%, so set max to 88%. -#define MIN_DIAL_IN_PUMP_PWM_DUTY_CYCLE 0.12 ///< Controller will error if PWM duty cycle < 10%, so set min to 12%. +#define MAX_DIAL_IN_PUMP_PWM_DUTY_CYCLE 0.89 ///< Controller will error if PWM duty cycle > 90%, so set max to 89%. +#define MIN_DIAL_IN_PUMP_PWM_DUTY_CYCLE 0.10 ///< Controller will error if PWM duty cycle < 10%, so set min to 10%. /// Interval (ms/task time) at which the dialIn pump is controlled. static const U32 DIP_CONTROL_INTERVAL = ( 10000 / TASK_GENERAL_INTERVAL ); @@ -83,14 +83,19 @@ #define DIP_MAX_CURR_WHEN_RUNNING_MA 2000.0 ///< Motor controller current should not exceed this when pump should be running. #define DIP_MAX_CURR_ERROR_DURATION_MS 2000 ///< Motor controller current errors persisting beyond this duration will trigger an alarm. -#define DIP_SPEED_ADC_TO_RPM_FACTOR 1.280938 ///< Conversion factor from ADC counts to RPM for dialIn pump motor. +#ifndef V2_0_SYSTEM + #define DIP_SPEED_ADC_TO_RPM_FACTOR 1.751752 ///< Conversion factor from ADC counts to RPM for dialIn pump motor. + #define DIP_MOTOR_RPM_TO_PWM_DC_FACTOR 0.00025 ///< ~40 BP motor RPM = 1% PWM duty cycle +#else + #define DIP_SPEED_ADC_TO_RPM_FACTOR 1.280938 ///< Conversion factor from ADC counts to RPM for blood pump motor + #define DIP_MOTOR_RPM_TO_PWM_DC_FACTOR 0.0003125 ///< ~32 BP motor RPM = 1% PWM duty cycle +#endif #define DIP_CURRENT_ADC_TO_MA_FACTOR 3.002 ///< Conversion factor from ADC counts to mA for dialIn pump motor. #define DIP_REV_PER_LITER 150.0 ///< Rotor revolutions per liter. /// Macro converts flow rate to motor RPM. #define DIP_ML_PER_MIN_TO_PUMP_RPM_FACTOR ( DIP_REV_PER_LITER / ML_PER_LITER ) #define DIP_GEAR_RATIO 32.0 ///< DialIn pump motor to dialIn pump gear ratio. -#define DIP_MOTOR_RPM_TO_PWM_DC_FACTOR 0.00028 ///< ~28 BP motor RPM = 1% PWM duty cycle. #define DIP_PWM_ZERO_OFFSET 0.1 ///< 10% PWM duty cycle = zero speed. /// Macro converts flow rate to estimate PWM needed to achieve it. #define DIP_PWM_FROM_ML_PER_MIN(rate) ( (rate) * DIP_ML_PER_MIN_TO_PUMP_RPM_FACTOR * DIP_GEAR_RATIO * DIP_MOTOR_RPM_TO_PWM_DC_FACTOR + DIP_PWM_ZERO_OFFSET )