Index: firmware/App/Controllers/DialOutFlow.c =================================================================== diff -u -r766708fceb0bdf1af8c7897df29d4f5036bfd3db -r6b870cd0699bb3ee22b93981d51373a6c2d56162 --- firmware/App/Controllers/DialOutFlow.c (.../DialOutFlow.c) (revision 766708fceb0bdf1af8c7897df29d4f5036bfd3db) +++ firmware/App/Controllers/DialOutFlow.c (.../DialOutFlow.c) (revision 6b870cd0699bb3ee22b93981d51373a6c2d56162) @@ -51,7 +51,8 @@ #define MAX_DIAL_OUT_PUMP_PWM_DUTY_CYCLE 0.88 ///< Controller will error if PWM duty cycle > 90%, so set max to 88%. #define MIN_DIAL_OUT_PUMP_PWM_DUTY_CYCLE 0.12 ///< Controller will error if PWM duty cycle < 10%, so set min to 12%. -#define DOP_CONTROL_INTERVAL ( 1000 / TASK_GENERAL_INTERVAL ) ///< Interval (ms/task time) at which the dialysate outlet pump is controlled. +/// Interval (ms/task time) at which the dialysate outlet pump is controlled. +static const U32 DOP_CONTROL_INTERVAL = ( 1000 / TASK_GENERAL_INTERVAL ); #define DOP_P_COEFFICIENT 0.0050 ///< P term for dialysate outlet pump control. #define DOP_I_COEFFICIENT 0.0001 ///< I term for dialysate outlet pump control. @@ -66,11 +67,16 @@ #define DOP_MAX_MOTOR_SPEED_WHILE_OFF_RPM 100.0 ///< Maximum motor speed (RPM) while motor is commanded off. #define DOP_MAX_ROTOR_VS_MOTOR_DIFF_RPM 5.0 ///< Maximum difference in speed between motor and rotor (in rotor RPM). #define DOP_MAX_MOTOR_SPEED_ERROR_RPM 300.0 ///< Maximum difference in speed between measured and commanded RPM. -#define DOP_OFF_ERROR_PERSIST ((5 * MS_PER_SECOND) / TASK_PRIORITY_INTERVAL) ///< Persist time (task intervals) for motor off error condition. -#define DOP_MOTOR_SPEED_ERROR_PERSIST ((5 * MS_PER_SECOND) / TASK_PRIORITY_INTERVAL) ///< Persist time (task intervals) motor speed error condition. -#define DOP_ROTOR_SPEED_ERROR_PERSIST ((12 * MS_PER_SECOND) / TASK_PRIORITY_INTERVAL) ///< Persist time (task intervals) rotor speed error condition. -#define DOP_DIRECTION_ERROR_PERSIST (250 / TASK_PRIORITY_INTERVAL) ///< Persist time (task intervals) pump direction error condition. +/// Persist time (task intervals) for motor off error condition. +static const U32 DOP_OFF_ERROR_PERSIST = ((5 * MS_PER_SECOND) / TASK_PRIORITY_INTERVAL); +/// Persist time (task intervals) motor speed error condition. +static const U32 DOP_MOTOR_SPEED_ERROR_PERSIST = ((5 * MS_PER_SECOND) / TASK_PRIORITY_INTERVAL); +/// Persist time (task intervals) rotor speed error condition. +static const U32 DOP_ROTOR_SPEED_ERROR_PERSIST = ((12 * MS_PER_SECOND) / TASK_PRIORITY_INTERVAL); +/// Persist time (task intervals) pump direction error condition. +static const U32 DOP_DIRECTION_ERROR_PERSIST = (250 / TASK_PRIORITY_INTERVAL); + #define DOP_MAX_CURR_WHEN_STOPPED_MA 150.0 ///< Motor controller current should not exceed this when pump should be stopped. #define DOP_MIN_CURR_WHEN_RUNNING_MA 150.0 ///< Motor controller current should always exceed this when pump should be running. #define DOP_MAX_CURR_WHEN_RUNNING_MA 2000.0 ///< Motor controller current should not exceed this when pump should be running.