Index: firmware/App/Controllers/DialOutFlow.c =================================================================== diff -u -r9034e1809869ded91cc2f878df0d9c0fd0b89d16 -rcbb126cfd3a02c08436a66ffecb2bdf926263705 --- firmware/App/Controllers/DialOutFlow.c (.../DialOutFlow.c) (revision 9034e1809869ded91cc2f878df0d9c0fd0b89d16) +++ firmware/App/Controllers/DialOutFlow.c (.../DialOutFlow.c) (revision cbb126cfd3a02c08436a66ffecb2bdf926263705) @@ -85,7 +85,6 @@ #define DOP_SPEED_ADC_TO_RPM_FACTOR 1.751752 ///< Conversion factor from ADC counts to RPM for dialysate outlet pump motor. #define DOP_MOTOR_RPM_TO_PWM_DC_FACTOR 0.000193 ///< ~52 BP motor RPM = 1% PWM duty cycle -#define UF_PWM_DC_PCT_PER_ML_PER_MIN 0.004 ///< Estimated DPo PWM duty cycle amount per each mL/min of ultrafiltration desired. #define DOP_CURRENT_ADC_TO_MA_FACTOR 3.002 ///< Conversion factor from ADC counts to mA for dialysate outlet pump motor. #define DOP_ADC_FULL_SCALE_V 3.0 ///< DPo analog signals are 0-3V (while int. ADC ref V may be different). #define DOP_ADC_ZERO 1998 ///< Mid-point (zero) for ADC readings.