Index: firmware/App/Controllers/PresOccl.c =================================================================== diff -u -r09e6cf9de34acf18f6e1138bf56ac0edb4821186 -r25a6eff69c9371f7a320751adf5359aaf070e5b0 --- firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 09e6cf9de34acf18f6e1138bf56ac0edb4821186) +++ firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 25a6eff69c9371f7a320751adf5359aaf070e5b0) @@ -363,15 +363,7 @@ // Check for stale arterial pressure reading if ( FALSE == isPersistentAlarmTriggered( ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR, ( lastArterialPressureReadCtr == artReadCtr || artErrorCtr > 0 ) ) ) { -#ifndef PBA_ESTIMATION arterialPressure.data = ARTERIAL_PRESSURE_V_PER_BIT * ( (F32)(artPres) / ( ARTERIAL_PRESSURE_SENSITIVITY * ARTERIAL_PRESSURE_V_BIAS ) ) + getF32OverrideValue( &arterialPressureOffset ); -#else - // TODO - temporary test code - remove later - F32 artPres = -0.1146 * getMeasuredBloodPumpMCSpeed(); - artPres = MAX(-200.0,artPres); - artPres = MIN(0,artPres); - arterialPressure.data = artPres; -#endif } else {