Index: firmware/App/Controllers/PresOccl.c =================================================================== diff -u -r25a6eff69c9371f7a320751adf5359aaf070e5b0 -r946f42bfd6606dd3d771d9a4a7b50e5678469f62 --- firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 25a6eff69c9371f7a320751adf5359aaf070e5b0) +++ firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 946f42bfd6606dd3d771d9a4a7b50e5678469f62) @@ -460,7 +460,7 @@ { F32 artPres = getFilteredArterialPressure(); -#ifndef DISABLE_PRESSURE_CHECKS +#ifndef DISABLE_ARTERIAL_PRESSURE_CHECK // Check arterial pressure is in range if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_HD_ARTERIAL_PRESSURE_OUT_OF_RANGE, ( artPres > ARTERIAL_PRESSURE_MAX_MMHG || artPres < ARTERIAL_PRESSURE_MIN_MMHG ) ) ) @@ -505,15 +505,15 @@ { F32 venPres = getFilteredVenousPressure(); -#ifndef DISABLE_PRESSURE_CHECKS - // Check arterial pressure is in range +#ifndef DISABLE_VENOUS_PRESSURE_CHECK + // Check venous pressure is in range if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_HD_VENOUS_PRESSURE_OUT_OF_RANGE, venPres > VENOUS_PRESSURE_MAX_MMHG ) || TRUE == isPersistentAlarmTriggered( ALARM_ID_HD_VENOUS_PRESSURE_OUT_OF_RANGE, venPres < VENOUS_PRESSURE_MIN_MMHG ) ) { SET_ALARM_WITH_1_F32_DATA( ALARM_ID_HD_VENOUS_PRESSURE_OUT_OF_RANGE, venPres ); } - // Check arterial pressure during treatment mode + // Check venous pressure during treatment mode if ( ( MODE_TREA == getCurrentOperationMode() ) && ( getTreatmentState() <= TREATMENT_DIALYSIS_STATE ) ) { F32 venLowLimit = (F32)getTreatmentParameterS32( TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT );