Index: firmware/App/Controllers/PresOccl.c =================================================================== diff -u -r37a9fd8f15e413db5337371a7d1a1cb65567af7c -r6eef422e39bed332f9958b165f8e64a23185a170 --- firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 37a9fd8f15e413db5337371a7d1a1cb65567af7c) +++ firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 6eef422e39bed332f9958b165f8e64a23185a170) @@ -669,7 +669,7 @@ static void checkOcclusions( void ) { U32 bpOccl = getMeasuredBloodPumpOcclusion(); - BOOL outOfRange = ( bpOccl < MIN_OCCLUSION_COUNTS || bpOccl > MAX_OCCLUSION_COUNTS ? TRUE : FALSE ); + BOOL outOfRange = ( bpOccl < MIN_OCCLUSION_COUNTS || bpOccl > MAX_OCCLUSION_COUNTS ? TRUE : FALSE ); if ( getSoftwareConfigStatus( SW_CONFIG_DISABLE_PRESSURE_CHECKS ) != SW_CONFIG_ENABLE_VALUE ) { @@ -679,14 +679,38 @@ SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_BP_OCCLUSION_OUT_OF_RANGE, bpOccl ); } - // Check for occlusion - if ( bpOccl > ( OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall ) ) + // Check for occlusion in PreTreatment modes when the cartridge is installed + // and setOcclusionInstallLevel has been called. + if( MODE_PRET == getCurrentOperationMode() ) { - signalBloodPumpHardStop(); // Stop pump immediately - SET_ALARM_WITH_1_U32_DATA( ALARM_ID_OCCLUSION_BLOOD_PUMP, bpOccl ) + if( getPreTreatmentSubState() > HD_PRE_TREATMENT_CART_INSTALL_STATE && getDrySelfTestsState() > DRY_SELF_TESTS_START_STATE ) + { + // Check for occlusion + if ( bpOccl > ( OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall ) ) + { + signalBloodPumpHardStop(); // Stop pump immediately + SET_ALARM_WITH_1_U32_DATA( ALARM_ID_OCCLUSION_BLOOD_PUMP, bpOccl ) + } + } } - else if ( bpOccl <= ( OCCLUSION_CLEAR_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall ) ) + + // Check for occlusion in PreTreatment modes where pumps are moving + else if( MODE_TREA == getCurrentOperationMode() ) { + if( TREATMENT_STOP_STATE != getTreatmentState()) + { + // Check for occlusion + if ( bpOccl > ( OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall ) ) + { + signalBloodPumpHardStop(); // Stop pump immediately + SET_ALARM_WITH_1_U32_DATA( ALARM_ID_OCCLUSION_BLOOD_PUMP, bpOccl ) + } + } + } + + // Clear alarm regardless of op mode once conditions are met. + if ( bpOccl <= ( OCCLUSION_CLEAR_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall ) ) + { clearAlarmCondition( ALARM_ID_OCCLUSION_BLOOD_PUMP ); } }