Index: firmware/App/Controllers/PresOccl.c =================================================================== diff -u -r6e23d627c3f5a1ef3f7cca5d4e48eb478b0eef4c -ra6d1ff9f7223af8efaa54f47c217d44391740405 --- firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 6e23d627c3f5a1ef3f7cca5d4e48eb478b0eef4c) +++ firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision a6d1ff9f7223af8efaa54f47c217d44391740405) @@ -86,7 +86,7 @@ #define ARTERIAL_PRESSURE_DIAG_CONDITION 3 ///< Arterial pressure status bits diagnostic condition (alarm). #define OCCLUSION_THRESHOLD_OFFSET 5500 ///< Threshold offset. Combined with initial reading after cartridge install, a threshold is derived above which an occlusion is detected. -#define OCCLUSION_CLEAR_THRESHOLD_OFFSET 5500 ///< Threshold offset. Combined with initial reading after cartridge install, a threshold is derived below which an occlusion is cleared. +#define OCCLUSION_CLEAR_THRESHOLD_OFFSET 5000 ///< Threshold offset. Combined with initial reading after cartridge install, a threshold is derived below which an occlusion is cleared. #define CARTRIDGE_LOADED_THRESHOLD 5000 ///< Threshold above which a cartridge is considered loaded. #define MIN_OCCLUSION_COUNTS 2000 ///< Minimum occlusion sensor reading for range check.