Index: firmware/App/Controllers/PresOccl.c =================================================================== diff -u -r7499a42cc0f906f9a4a947c82c5b4615217ce7e5 -r818d96fff963c707f5e2cb5aae7bf16f23e4b08a --- firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 7499a42cc0f906f9a4a947c82c5b4615217ce7e5) +++ firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 818d96fff963c707f5e2cb5aae7bf16f23e4b08a) @@ -22,7 +22,8 @@ #include "NVDataMgmt.h" #include "OperationModes.h" #include "PersistentAlarm.h" -#include "SystemCommMessages.h" +#include "SystemCommMessages.h" +#include "SelfTests.h" #include "TaskGeneral.h" #include "Temperatures.h" #include "Timers.h" @@ -709,7 +710,7 @@ static void checkOcclusions( void ) { U32 bpOccl = getMeasuredBloodPumpOcclusion(); - BOOL outOfRange = ( bpOccl < MIN_OCCLUSION_COUNTS || bpOccl > MAX_OCCLUSION_COUNTS ? TRUE : FALSE ); + BOOL outOfRange = ( bpOccl < MIN_OCCLUSION_COUNTS || bpOccl > MAX_OCCLUSION_COUNTS ? TRUE : FALSE ); #ifndef _RELEASE_ if ( getSoftwareConfigStatus( SW_CONFIG_DISABLE_PRESSURE_CHECKS ) != SW_CONFIG_ENABLE_VALUE ) @@ -721,14 +722,41 @@ SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_BP_OCCLUSION_OUT_OF_RANGE, bpOccl ); } - // Check for occlusion - if ( bpOccl > ( OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall ) ) + // Check for occlusion in PreTreatment modes when the cartridge is installed + // and setOcclusionInstallLevel has been called. + if( MODE_PRET == getCurrentOperationMode() ) { - signalBloodPumpHardStop(); // Stop pump immediately - SET_ALARM_WITH_1_U32_DATA( ALARM_ID_OCCLUSION_BLOOD_PUMP, bpOccl ) + + if( getPreTreatmentSubState() > HD_PRE_TREATMENT_CART_INSTALL_STATE && getDrySelfTestsState() > DRY_SELF_TESTS_START_STATE ) + { + // Check for occlusion + if ( bpOccl > ( OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall ) ) + { + signalBloodPumpHardStop(); // Stop pump immediately + SET_ALARM_WITH_1_U32_DATA( ALARM_ID_OCCLUSION_BLOOD_PUMP, bpOccl ) + } + } } - else + + // Check for occlusion in PreTreatment modes where pumps are moving + else if( MODE_TREA == getCurrentOperationMode() ) { + + if( TREATMENT_STOP_STATE != getTreatmentState()) + { + + // Check for occlusion + if ( bpOccl > ( OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall ) ) + { + signalBloodPumpHardStop(); // Stop pump immediately + SET_ALARM_WITH_1_U32_DATA( ALARM_ID_OCCLUSION_BLOOD_PUMP, bpOccl ) + } + } + } + + // Clear alarm regardless of op mode once conditions are met. + if ( bpOccl <= ( OCCLUSION_CLEAR_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall ) ) + { clearAlarmCondition( ALARM_ID_OCCLUSION_BLOOD_PUMP ); } }