Index: firmware/App/Controllers/PresOccl.c =================================================================== diff -u -r8cebc7f282f403c99f712d422454c15414b6fc73 -r84e94699ac1a6236a6aba130e45149a9cf3ac5f8 --- firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 8cebc7f282f403c99f712d422454c15414b6fc73) +++ firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 84e94699ac1a6236a6aba130e45149a9cf3ac5f8) @@ -363,7 +363,15 @@ // Check for stale arterial pressure reading if ( FALSE == isPersistentAlarmTriggered( ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR, ( lastArterialPressureReadCtr == artReadCtr || artErrorCtr > 0 ) ) ) { +#ifndef PBA_ESTIMATION arterialPressure.data = ARTERIAL_PRESSURE_V_PER_BIT * ( (F32)(artPres) / ( ARTERIAL_PRESSURE_SENSITIVITY * ARTERIAL_PRESSURE_V_BIAS ) ) + getF32OverrideValue( &arterialPressureOffset ); +#else + // TODO - temporary test code - remove later + F32 artPres = -0.1146 * getMeasuredBloodPumpMCSpeed(); + artPres = MAX(-200.0,artPres); + artPres = MIN(0,artPres); + arterialPressure.data = artPres; +#endif } else {