Index: firmware/App/Controllers/PresOccl.c =================================================================== diff -u -r933a18d740285e70be9d00696ed0f5a5381bc8e4 -r40a0bc77c53a4eb538580d44f1117d26f1d37fb2 --- firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 933a18d740285e70be9d00696ed0f5a5381bc8e4) +++ firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 40a0bc77c53a4eb538580d44f1117d26f1d37fb2) @@ -56,7 +56,7 @@ #define OCCLUSION_THRESHOLD 25000 ///< Threshold above which an occlusion is detected. #define CARTRIDGE_LOADED_THRESHOLD 5000 ///< Threshold above which a cartridge is considered loaded. -#define PRES_ALARM_PERSISTENCE ( MS_PER_SECOND / TASK_GENERAL_INTERVAL ) ///< Alarm persistence period for pressure alarms. +#define PRES_ALARM_PERSISTENCE ( 1 * MS_PER_SECOND ) ///< Alarm persistence period for pressure alarms. /// Defined states for the pressure and occlusion monitor state machine. typedef enum PresOccl_States