Index: firmware/App/Controllers/PresOccl.c =================================================================== diff -u -r9bb662bfd053fa114ac47224db7825c85bb6d796 -r7dc4bb045389ac95e6658602f2659f6071542361 --- firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 9bb662bfd053fa114ac47224db7825c85bb6d796) +++ firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 7dc4bb045389ac95e6658602f2659f6071542361) @@ -191,10 +191,10 @@ void initPresOccl( void ) { // Initialize persistent pressure alarms - initPersistentAlarm( ALARM_ID_ARTERIAL_PRESSURE_LOW, 0, PRES_ALARM_PERSISTENCE ); - initPersistentAlarm( ALARM_ID_ARTERIAL_PRESSURE_HIGH, 0, PRES_ALARM_PERSISTENCE ); - initPersistentAlarm( ALARM_ID_VENOUS_PRESSURE_LOW, 0, PRES_ALARM_PERSISTENCE ); - initPersistentAlarm( ALARM_ID_VENOUS_PRESSURE_HIGH, 0, PRES_ALARM_PERSISTENCE ); + initPersistentAlarm( ALARM_ID_HD_ARTERIAL_PRESSURE_LOW, 0, PRES_ALARM_PERSISTENCE ); + initPersistentAlarm( ALARM_ID_HD_ARTERIAL_PRESSURE_HIGH, 0, PRES_ALARM_PERSISTENCE ); + initPersistentAlarm( ALARM_ID_HD_VENOUS_PRESSURE_LOW, 0, PRES_ALARM_PERSISTENCE ); + initPersistentAlarm( ALARM_ID_HD_VENOUS_PRESSURE_HIGH, 0, PRES_ALARM_PERSISTENCE ); initPersistentAlarm( ALARM_ID_HD_ARTERIAL_PRESSURE_OUT_OF_RANGE, 0, PRES_ALARM_PERSISTENCE ); initPersistentAlarm( ALARM_ID_HD_VENOUS_PRESSURE_OUT_OF_RANGE, 0, PRES_ALARM_PERSISTENCE ); initPersistentAlarm( ALARM_ID_HD_BP_OCCLUSION_OUT_OF_RANGE, 0, PRES_OCCL_ALARM_PERSISTENCE ); @@ -503,14 +503,14 @@ artPresHigh = ( artPres > artHighLimit ? TRUE : FALSE ); // Check arterial pressure is within user set alarm limits - if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_ARTERIAL_PRESSURE_LOW, artPresLow ) ) + if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_HD_ARTERIAL_PRESSURE_LOW, artPresLow ) ) { - SET_ALARM_WITH_2_F32_DATA( ALARM_ID_ARTERIAL_PRESSURE_LOW, artPres, artLowLimit ); + SET_ALARM_WITH_2_F32_DATA( ALARM_ID_HD_ARTERIAL_PRESSURE_LOW, artPres, artLowLimit ); } - if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_ARTERIAL_PRESSURE_HIGH, artPresHigh ) ) + if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_HD_ARTERIAL_PRESSURE_HIGH, artPresHigh ) ) { - SET_ALARM_WITH_2_F32_DATA( ALARM_ID_ARTERIAL_PRESSURE_HIGH, artPres, artHighLimit ); + SET_ALARM_WITH_2_F32_DATA( ALARM_ID_HD_ARTERIAL_PRESSURE_HIGH, artPres, artHighLimit ); } } else if ( MODE_TREA == getCurrentOperationMode() ) // All treatment states not covered by the if above @@ -521,19 +521,19 @@ BOOL artPresHigh = ( artPres > artHighLimit ? TRUE : FALSE ); // Check arterial pressure is within max selectable range - if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_ARTERIAL_PRESSURE_LOW, artPresLow ) ) + if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_HD_ARTERIAL_PRESSURE_LOW, artPresLow ) ) { - SET_ALARM_WITH_2_F32_DATA( ALARM_ID_ARTERIAL_PRESSURE_LOW, artPres, artLowLimit ); + SET_ALARM_WITH_2_F32_DATA( ALARM_ID_HD_ARTERIAL_PRESSURE_LOW, artPres, artLowLimit ); } - if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_ARTERIAL_PRESSURE_HIGH, artPresHigh ) ) + if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_HD_ARTERIAL_PRESSURE_HIGH, artPresHigh ) ) { - SET_ALARM_WITH_2_F32_DATA( ALARM_ID_ARTERIAL_PRESSURE_HIGH, artPres, artHighLimit ); + SET_ALARM_WITH_2_F32_DATA( ALARM_ID_HD_ARTERIAL_PRESSURE_HIGH, artPres, artHighLimit ); } } else { // Reset persistence if alarm is out of scope - isPersistentAlarmTriggered( ALARM_ID_ARTERIAL_PRESSURE_LOW, FALSE ); - isPersistentAlarmTriggered( ALARM_ID_ARTERIAL_PRESSURE_HIGH, FALSE ); + isPersistentAlarmTriggered( ALARM_ID_HD_ARTERIAL_PRESSURE_LOW, FALSE ); + isPersistentAlarmTriggered( ALARM_ID_HD_ARTERIAL_PRESSURE_HIGH, FALSE ); } } } @@ -584,19 +584,19 @@ // Cannot monitor for low venous pressure while venting air trap if ( getValveAirTrapStatus() != STATE_OPEN ) { - if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_VENOUS_PRESSURE_LOW, venPresLow ) ) + if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_HD_VENOUS_PRESSURE_LOW, venPresLow ) ) { - SET_ALARM_WITH_2_F32_DATA( ALARM_ID_VENOUS_PRESSURE_LOW, venPres, venLowLimit ); + SET_ALARM_WITH_2_F32_DATA( ALARM_ID_HD_VENOUS_PRESSURE_LOW, venPres, venLowLimit ); } } else { // clear persistence if air trap valve is open - isPersistentAlarmTriggered( ALARM_ID_VENOUS_PRESSURE_LOW, FALSE ); + isPersistentAlarmTriggered( ALARM_ID_HD_VENOUS_PRESSURE_LOW, FALSE ); } - if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_VENOUS_PRESSURE_HIGH, venPresHigh ) ) + if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_HD_VENOUS_PRESSURE_HIGH, venPresHigh ) ) { - SET_ALARM_WITH_2_F32_DATA( ALARM_ID_VENOUS_PRESSURE_HIGH, venPres, venHighLimit ); + SET_ALARM_WITH_2_F32_DATA( ALARM_ID_HD_VENOUS_PRESSURE_HIGH, venPres, venHighLimit ); } } else if ( MODE_TREA == getCurrentOperationMode() ) // All treatment states not covered by the if statement above @@ -607,19 +607,19 @@ BOOL venPresHigh = ( venPres > venHighLimit ? TRUE : FALSE ); // Check venous pressure is within max selectable range - if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_VENOUS_PRESSURE_HIGH, venPresHigh ) ) + if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_HD_VENOUS_PRESSURE_HIGH, venPresHigh ) ) { - SET_ALARM_WITH_2_F32_DATA( ALARM_ID_VENOUS_PRESSURE_HIGH, venPres, venHighLimit ); + SET_ALARM_WITH_2_F32_DATA( ALARM_ID_HD_VENOUS_PRESSURE_HIGH, venPres, venHighLimit ); } - if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_VENOUS_PRESSURE_LOW, venPresLow ) ) + if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_HD_VENOUS_PRESSURE_LOW, venPresLow ) ) { - SET_ALARM_WITH_2_F32_DATA( ALARM_ID_VENOUS_PRESSURE_LOW, venPres, venLowLimit ); + SET_ALARM_WITH_2_F32_DATA( ALARM_ID_HD_VENOUS_PRESSURE_LOW, venPres, venLowLimit ); } } else { // Reset persistence if alarm is out of scope - isPersistentAlarmTriggered( ALARM_ID_VENOUS_PRESSURE_LOW, FALSE ); - isPersistentAlarmTriggered( ALARM_ID_VENOUS_PRESSURE_HIGH, FALSE ); + isPersistentAlarmTriggered( ALARM_ID_HD_VENOUS_PRESSURE_LOW, FALSE ); + isPersistentAlarmTriggered( ALARM_ID_HD_VENOUS_PRESSURE_HIGH, FALSE ); } } } @@ -671,7 +671,7 @@ if ( bpOccl > ( OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall ) ) { signalBloodPumpHardStop(); // Stop pump immediately - SET_ALARM_WITH_2_U32_DATA( ALARM_ID_OCCLUSION_BLOOD_PUMP, bpOccl, OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall ) + SET_ALARM_WITH_2_U32_DATA( ALARM_ID_HD_OCCLUSION_BLOOD_PUMP, bpOccl, OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall ) } } } @@ -685,15 +685,15 @@ if ( bpOccl > ( OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall ) ) { signalBloodPumpHardStop(); // Stop pump immediately - SET_ALARM_WITH_2_U32_DATA( ALARM_ID_OCCLUSION_BLOOD_PUMP, bpOccl, OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall ) + SET_ALARM_WITH_2_U32_DATA( ALARM_ID_HD_OCCLUSION_BLOOD_PUMP, bpOccl, OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall ) } } } // Clear alarm regardless of op mode once conditions are met. if ( bpOccl <= ( OCCLUSION_CLEAR_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall ) ) { - clearAlarmCondition( ALARM_ID_OCCLUSION_BLOOD_PUMP ); + clearAlarmCondition( ALARM_ID_HD_OCCLUSION_BLOOD_PUMP ); } // Check for venous occlusion @@ -938,7 +938,7 @@ if ( ( bpPressure < CARTRIDGE_LOADED_THRESHOLD ) || ( bpPressure > OCCLUSION_CARTRIDGE_LOADED_PRESSURE_READING_MAX ) ) { - activateAlarmNoData( ALARM_ID_NO_CARTRIDGE_LOADED ); + activateAlarmNoData( ALARM_ID_HD_NO_CARTRIDGE_LOADED ); result = SELF_TEST_STATUS_FAILED; } }