Index: firmware/App/Controllers/PresOccl.c =================================================================== diff -u -rc71ff8d9b1c3efea8649e11d12da472f06a5a511 -r8373758a03ea82b46b4b313d2f2b24e3974a17f6 --- firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision c71ff8d9b1c3efea8649e11d12da472f06a5a511) +++ firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 8373758a03ea82b46b4b313d2f2b24e3974a17f6) @@ -59,13 +59,11 @@ #define ARTERIAL_PRESSURE_SELF_TEST_MAX ( 100.0F ) ///< Maximum self-test value for arterial pressure sensor reading (in mmHg). #define ARTERIAL_PRESSURE_MAX_MMHG ( 2000.0F ) ///< Maximum arterial pressure reading (in mmHg) for range check. #define ARTERIAL_PRESSURE_MIN_MMHG ( -1500.0F ) ///< Minimum arterial pressure reading (in mmHg) for range check. -#define ARTERIAL_PRESSURE_ND_MIN_MMHG ( -250.0F ) ///< Minimum arterial pressure reading (in mmHg) when not dialyzing. #define VENOUS_PRESSURE_SELF_TEST_MIN ( -100.0F ) ///< Minimum self-test value for venous pressure sensor reading (in mmHg). #define VENOUS_PRESSURE_SELF_TEST_MAX ( 600.0F ) ///< Maximum self-test value for venous pressure sensor reading (in mmHg). #define VENOUS_PRESSURE_MAX_MMHG ( 2000.0F ) ///< Maximum venous pressure reading (in mmHg) for range check. #define VENOUS_PRESSURE_MIN_MMHG ( -1500.0F ) ///< Minimum venous pressure reading (in mmHg) for range check. -#define VENOUS_PRESSURE_ND_MAX_MMHG ( 300.0F ) ///< Maximum venous pressure reading (in mmHg) when not dialyzing. #define PSI_TO_MMHG ( 51.7149F ) ///< Conversion factor for converting PSI to mmHg. @@ -587,8 +585,7 @@ } // Check arterial pressure during treatment mode - if ( ( MODE_TREA == getCurrentOperationMode() ) && - ( getTreatmentState() <= TREATMENT_DIALYSIS_STATE && getDialysisState() != DIALYSIS_SALINE_BOLUS_STATE ) ) + if ( ( MODE_TREA == getCurrentOperationMode() ) && ( getTreatmentState() <= TREATMENT_DIALYSIS_STATE ) ) { F32 artLowLimit = (F32)getTreatmentParameterS32( TREATMENT_PARAM_ART_PRESSURE_LOW_LIMIT ); F32 artHighLimit = (F32)getTreatmentParameterS32( TREATMENT_PARAM_ART_PRESSURE_HIGH_LIMIT ); @@ -611,16 +608,6 @@ SET_ALARM_WITH_2_F32_DATA( ALARM_ID_ARTERIAL_PRESSURE_HIGH, artPres, artHighLimit ); } } - else if (( MODE_TREA == getCurrentOperationMode() ) && ( TREATMENT_STOP_STATE == getTreatmentState() || - TREATMENT_RINSEBACK_STATE == getTreatmentState() || TREATMENT_RECIRC_STATE == getTreatmentState() || - TREATMENT_END_STATE == getTreatmentState())) - { - // Check arterial pressure is within non-treatment alarm limits - if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_ARTERIAL_PRESSURE_LOW, artPres < ARTERIAL_PRESSURE_ND_MIN_MMHG ) ) - { - SET_ALARM_WITH_2_F32_DATA( ALARM_ID_ARTERIAL_PRESSURE_LOW, artPres, ARTERIAL_PRESSURE_ND_MIN_MMHG ); - } - } else { // Reset persistence if alarm is out of scope isPersistentAlarmTriggered( ALARM_ID_ARTERIAL_PRESSURE_LOW, FALSE ); @@ -653,8 +640,7 @@ } // Check venous pressure during treatment mode - if ( ( MODE_TREA == getCurrentOperationMode() ) && - ( getTreatmentState() <= TREATMENT_DIALYSIS_STATE && getDialysisState() != DIALYSIS_SALINE_BOLUS_STATE ) ) + if ( ( MODE_TREA == getCurrentOperationMode() ) && ( getTreatmentState() <= TREATMENT_DIALYSIS_STATE ) ) { F32 venLowLimit = (F32)getTreatmentParameterS32( TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT ); F32 venHighLimit = (F32)getTreatmentParameterS32( TREATMENT_PARAM_VEN_PRESSURE_HIGH_LIMIT ); @@ -684,16 +670,6 @@ SET_ALARM_WITH_2_F32_DATA( ALARM_ID_VENOUS_PRESSURE_HIGH, venPres, venHighLimit ); } } - else if ( ( MODE_TREA == getCurrentOperationMode() ) && ( TREATMENT_STOP_STATE == getTreatmentState() || - TREATMENT_RINSEBACK_STATE == getTreatmentState() || TREATMENT_RECIRC_STATE == getTreatmentState() || - TREATMENT_END_STATE == getTreatmentState())) - { - // Check venous pressure is less than non-treatment alarm limit - if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_VENOUS_PRESSURE_HIGH, venPres > VENOUS_PRESSURE_ND_MAX_MMHG ) ) - { - SET_ALARM_WITH_2_F32_DATA( ALARM_ID_VENOUS_PRESSURE_HIGH, venPres, VENOUS_PRESSURE_ND_MAX_MMHG ); - } - } else { // Reset persistence if alarm is out of scope isPersistentAlarmTriggered( ALARM_ID_VENOUS_PRESSURE_LOW, FALSE );