Index: firmware/App/Controllers/PresOccl.c =================================================================== diff -u -rd0e3bb3d509ad4583a821c9448eda0c908f3dc63 -r80dc7bfd2d00d032ae0e2f72830315404e9b9dca --- firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision d0e3bb3d509ad4583a821c9448eda0c908f3dc63) +++ firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 80dc7bfd2d00d032ae0e2f72830315404e9b9dca) @@ -97,7 +97,6 @@ #define CARTRIDGE_LOADED_THRESHOLD 5000 ///< Threshold above which a cartridge is considered loaded. #define MIN_OCCLUSION_COUNTS 2000 ///< Minimum occlusion sensor reading for range check. -#define MAX_OCCLUSION_COUNTS 32766 ///< Maximum occlusion sensor reading for range check. #define OCCLUSION_CARTRIDGE_LOADED_PRESSURE_READING_MAX 24500 ///< Occlusion sensors maximum pressure reading limit when cartridge is considered loaded. #define EMPTY_SALINE_BAG_THRESHOLD_MMHG -300.0F ///< Threshold below which the saline bag is considered empty (in mmHg). @@ -790,7 +789,7 @@ static void checkOcclusions( void ) { U32 bpOccl = getMeasuredBloodPumpOcclusion(); - BOOL outOfRange = ( bpOccl < MIN_OCCLUSION_COUNTS || bpOccl > MAX_OCCLUSION_COUNTS ? TRUE : FALSE ); + BOOL outOfRange = ( bpOccl < MIN_OCCLUSION_COUNTS ? TRUE : FALSE ); #ifndef _RELEASE_ if( HW_CONFIG_BETA == getHardwareConfigStatus() )