Index: firmware/App/Controllers/PresOccl.c =================================================================== diff -u -rd91a24c730aeb5cd7e3eba9ef4eca78e442911f8 -r6d044be7a5c4ef961c2d2569d88be95c9bd86be6 --- firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision d91a24c730aeb5cd7e3eba9ef4eca78e442911f8) +++ firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 6d044be7a5c4ef961c2d2569d88be95c9bd86be6) @@ -64,8 +64,6 @@ static PRESSURE_STATE_T presOcclState = PRESSURE_INIT_STATE; ///< current state of pressure monitor state machine. static U32 presOcclDataPublicationTimerCounter = 0; ///< used to schedule pressure data publication to CAN bus. -static OVERRIDE_F32_T measuredBloodFlowRate = { 0.0, 0.0, 0.0, 0 }; ///< measured blood flow rate - static OVERRIDE_U32_T presOcclDataPublishInterval = { PRES_OCCL_DATA_PUB_INTERVAL, PRES_OCCL_DATA_PUB_INTERVAL, 0, 0 }; ///< interval (in ms) at which to publish pressure/occlusion data to CAN bus. static OVERRIDE_F32_T arterialPressure = {0.0, 0.0, 0.0, 0 }; ///< measured arterial pressure. static OVERRIDE_F32_T venousPressure = {0.0, 0.0, 0.0, 0 }; ///< measured venous pressure. @@ -74,7 +72,7 @@ static OVERRIDE_F32_T dialOutPumpOcclusion = {0.0, 0.0, 0.0, 0 }; ///< measured dialysate outlet pump occlusion pressure. static PRESSURE_SELF_TEST_STATE_T presOcclSelfTestState = PRESSURE_SELF_TEST_STATE_START; ///< current pressure self-test state. -static U32 bloodPumpSelfTestTimerCount = 0; ///< timer counter for pressure self-test. +static U32 presOcclSelfTestTimerCount = 0; ///< timer counter for pressure self-test. static F32 arterialPressureLowLimitmmHG = 0.0; ///< lower alarm limit for arterial pressure. static F32 arterialPressureHighLimitmmHG = 0.0; ///< upper alarm limit for arterial pressure.