Index: firmware/App/Controllers/PresOccl.c =================================================================== diff -u -rdc8530d1b15e4b1199cacf0284dbe8f7802e9dde -r818d96fff963c707f5e2cb5aae7bf16f23e4b08a --- firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision dc8530d1b15e4b1199cacf0284dbe8f7802e9dde) +++ firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 818d96fff963c707f5e2cb5aae7bf16f23e4b08a) @@ -8,7 +8,7 @@ * @file PresOccl.c * * @author (last) Dara Navaei -* @date (last) 26-May-2022 +* @date (last) 13-Jul-2022 * * @author (original) Sean * @date (original) 15-Jan-2020 @@ -60,11 +60,13 @@ #define ARTERIAL_PRESSURE_SELF_TEST_MAX ( 100.0F ) ///< Maximum self-test value for arterial pressure sensor reading (in mmHg). #define ARTERIAL_PRESSURE_MAX_MMHG ( 2000.0F ) ///< Maximum arterial pressure reading (in mmHg) for range check. #define ARTERIAL_PRESSURE_MIN_MMHG ( -1500.0F ) ///< Minimum arterial pressure reading (in mmHg) for range check. +#define ARTERIAL_PRESSURE_ND_MIN_MMHG ( -250.0F ) ///< Minimum arterial pressure reading (in mmHg) when not dialyzing. #define VENOUS_PRESSURE_SELF_TEST_MIN ( -100.0F ) ///< Minimum self-test value for venous pressure sensor reading (in mmHg). #define VENOUS_PRESSURE_SELF_TEST_MAX ( 600.0F ) ///< Maximum self-test value for venous pressure sensor reading (in mmHg). #define VENOUS_PRESSURE_MAX_MMHG ( 2000.0F ) ///< Maximum venous pressure reading (in mmHg) for range check. #define VENOUS_PRESSURE_MIN_MMHG ( -1500.0F ) ///< Minimum venous pressure reading (in mmHg) for range check. +#define VENOUS_PRESSURE_ND_MAX_MMHG ( 300.0F ) ///< Maximum venous pressure reading (in mmHg) when not dialyzing. #define PSI_TO_MMHG ( 51.7149F ) ///< Conversion factor for converting PSI to mmHg. @@ -586,7 +588,8 @@ } // Check arterial pressure during treatment mode - if ( ( MODE_TREA == getCurrentOperationMode() ) && ( getTreatmentState() <= TREATMENT_DIALYSIS_STATE ) ) + if ( ( MODE_TREA == getCurrentOperationMode() ) && + ( TREATMENT_DIALYSIS_STATE == getTreatmentState() ) && ( getDialysisState() != DIALYSIS_SALINE_BOLUS_STATE ) ) { F32 artLowLimit = (F32)getTreatmentParameterS32( TREATMENT_PARAM_ART_PRESSURE_LOW_LIMIT ); F32 artHighLimit = (F32)getTreatmentParameterS32( TREATMENT_PARAM_ART_PRESSURE_HIGH_LIMIT ); @@ -609,6 +612,14 @@ SET_ALARM_WITH_2_F32_DATA( ALARM_ID_ARTERIAL_PRESSURE_HIGH, artPres, artHighLimit ); } } + else if ( MODE_TREA == getCurrentOperationMode() ) // All treatment states not covered by the if above + { + // Check arterial pressure is within non-treatment alarm limits + if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_ARTERIAL_PRESSURE_LOW, artPres < ARTERIAL_PRESSURE_ND_MIN_MMHG ) ) + { + SET_ALARM_WITH_2_F32_DATA( ALARM_ID_ARTERIAL_PRESSURE_LOW, artPres, ARTERIAL_PRESSURE_ND_MIN_MMHG ); + } + } else { // Reset persistence if alarm is out of scope isPersistentAlarmTriggered( ALARM_ID_ARTERIAL_PRESSURE_LOW, FALSE ); @@ -641,7 +652,8 @@ } // Check venous pressure during treatment mode - if ( ( MODE_TREA == getCurrentOperationMode() ) && ( getTreatmentState() <= TREATMENT_DIALYSIS_STATE ) ) + if ( MODE_TREA == getCurrentOperationMode() && + TREATMENT_DIALYSIS_STATE == getTreatmentState() && getDialysisState() != DIALYSIS_SALINE_BOLUS_STATE ) { F32 venLowLimit = (F32)getTreatmentParameterS32( TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT ); F32 venHighLimit = (F32)getTreatmentParameterS32( TREATMENT_PARAM_VEN_PRESSURE_HIGH_LIMIT ); @@ -671,6 +683,14 @@ SET_ALARM_WITH_2_F32_DATA( ALARM_ID_VENOUS_PRESSURE_HIGH, venPres, venHighLimit ); } } + else if ( MODE_TREA == getCurrentOperationMode() ) // All treatment states not covered by the if statement above + { + // Check venous pressure is less than non-treatment alarm limit + if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_VENOUS_PRESSURE_HIGH, venPres > VENOUS_PRESSURE_ND_MAX_MMHG ) ) + { + SET_ALARM_WITH_2_F32_DATA( ALARM_ID_VENOUS_PRESSURE_HIGH, venPres, VENOUS_PRESSURE_ND_MAX_MMHG ); + } + } else { // Reset persistence if alarm is out of scope isPersistentAlarmTriggered( ALARM_ID_VENOUS_PRESSURE_LOW, FALSE );