Index: firmware/App/Controllers/PresOccl.c =================================================================== diff -u -rdf4242fcc91f8cd1aada438b33fa845c223d0e0e -r8d829681af297f6638bc7b0933094305fc9f2b8e --- firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision df4242fcc91f8cd1aada438b33fa845c223d0e0e) +++ firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 8d829681af297f6638bc7b0933094305fc9f2b8e) @@ -712,6 +712,15 @@ U32 bpOccl = getMeasuredBloodPumpOcclusion(); BOOL outOfRange = ( bpOccl < MIN_OCCLUSION_COUNTS || bpOccl > MAX_OCCLUSION_COUNTS ? TRUE : FALSE ); +#ifndef _RELEASE + if( getSoftwareConfigStatus( SW_CONFIG_ENABLE_V3_SYSTEM ) == SW_CONFIG_ENABLE_VALUE) + { + U32 minOcclusionCountV3 = 1500; + outOfRange = (( bpOccl < minOcclusionCountV3 || bpOccl > MAX_OCCLUSION_COUNTS ? TRUE : FALSE )); + } +#endif + + #ifndef _RELEASE_ if ( getSoftwareConfigStatus( SW_CONFIG_DISABLE_PRESSURE_CHECKS ) != SW_CONFIG_ENABLE_VALUE ) #endif