Index: firmware/App/Controllers/PresOccl.c =================================================================== diff -u -rf6b78d1fe6741043de38707211710ab0e8a08483 -r23016be42e59935da5aa53b26b8cb31e2a240858 --- firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision f6b78d1fe6741043de38707211710ab0e8a08483) +++ firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 23016be42e59935da5aa53b26b8cb31e2a240858) @@ -78,8 +78,8 @@ #define ARTERIAL_PRESSURE_STALE_DATA 2 ///< Arterial pressure status bits indicate data is stale (no new data since last fpga read). #define ARTERIAL_PRESSURE_DIAG_CONDITION 3 ///< Arterial pressure status bits diagnostic condition (alarm). -#define OCCLUSION_THRESHOLD_OFFSET 10000 ///< Threshold offset. Combined with initial reading after cartridge install, a threshold is derived above which an occlusion is detected. -#define OCCLUSION_CLEAR_THRESHOLD_OFFSET 6000 ///< Threshold offset. Combined with initial reading after cartridge install, a threshold is derived below which an occlusion is cleared. +#define OCCLUSION_THRESHOLD_OFFSET 4000 ///< Threshold offset. Combined with initial reading after cartridge install, a threshold is derived above which an occlusion is detected. +#define OCCLUSION_CLEAR_THRESHOLD_OFFSET 4000 ///< Threshold offset. Combined with initial reading after cartridge install, a threshold is derived below which an occlusion is cleared. #define CARTRIDGE_LOADED_THRESHOLD 5000 ///< Threshold above which a cartridge is considered loaded. #define MIN_OCCLUSION_COUNTS 2000 ///< Minimum occlusion sensor reading for range check.