Index: firmware/App/Controllers/SyringePump.c =================================================================== diff -u -r3d72b777cf1ceb673d118341c46e2d6d5b7b75f5 -rf3c83b15c8fc871808fef3ae236189b844ecbcb8 --- firmware/App/Controllers/SyringePump.c (.../SyringePump.c) (revision 3d72b777cf1ceb673d118341c46e2d6d5b7b75f5) +++ firmware/App/Controllers/SyringePump.c (.../SyringePump.c) (revision f3c83b15c8fc871808fef3ae236189b844ecbcb8) @@ -59,7 +59,7 @@ #define MIN_HEPARIN_CONTINUOUS_RATE 0.2 ///< Minimum continuous Heparin flow rate (in mL/hr). #define MAX_HEPARIN_CONTINUOUS_RATE 1.0 ///< Maximum continuous Heparin flow rate (in mL/hr). -#define HEPARIN_BOLUS_TARGET_RATE 40.0 ///< Heparin bolus target rate (in ml/hr); +#define HEPARIN_BOLUS_TARGET_RATE 720.0F ///< Heparin bolus target rate (in ml/hr); #define SYRINGE_PUMP_RETRACT_RATE 3600.0 ///< Retract rate is 5 mL/ 5 s = 1 mL/s = 3,600 mL/hr. #define SYRINGE_PUMP_SEEK_RATE 3600.0 ///< Seek plunger rate is 5 mL/ 5 s = 1 mL/s = 3,600 mL/hr. #define SYRINGE_PUMP_PRIME_RATE 635.0 ///< Prime rate is 0.5 mm ^ 2 x PI x 450 mm = 0.353 mL / 2s = 635 mL/hr.