Index: firmware/App/Controllers/SyringePump.c =================================================================== diff -u -r4f89cf6acdca1a537cd9d9bd851dc9dc7cf5c63a -rd75fca5ec82e1dfe92df0c712b21dc3a98829d0e --- firmware/App/Controllers/SyringePump.c (.../SyringePump.c) (revision 4f89cf6acdca1a537cd9d9bd851dc9dc7cf5c63a) +++ firmware/App/Controllers/SyringePump.c (.../SyringePump.c) (revision d75fca5ec82e1dfe92df0c712b21dc3a98829d0e) @@ -7,8 +7,8 @@ * * @file SyringePump.c * -* @author (last) Darren Cox -* @date (last) 09-Oct-2023 +* @author (last) Sean Nash +* @date (last) 10-Oct-2023 * * @author (original) Sean Nash * @date (original) 04-Mar-2021 @@ -1111,7 +1111,7 @@ clearAlarmCondition( ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED ); } // If no syringe detected and detection is required, trigger alarm condition - else if ( ( FALSE == isSyringeDetected() ) && ( TRUE == requireSyringeDetection ) ) + else if ( FALSE == isSyringeDetected() ) { activateAlarmNoData( ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED ); }