Index: firmware/App/Controllers/SyringePump.c =================================================================== diff -u -r6f1e959a3c5f991d7b8a787939d252c7998cbbd1 -rf499e4d718aaf99432e4ed4300a0b48e402feb60 --- firmware/App/Controllers/SyringePump.c (.../SyringePump.c) (revision 6f1e959a3c5f991d7b8a787939d252c7998cbbd1) +++ firmware/App/Controllers/SyringePump.c (.../SyringePump.c) (revision f499e4d718aaf99432e4ed4300a0b48e402feb60) @@ -1066,12 +1066,12 @@ else if ( syringePumpControllerMeasuredDirection != MOTOR_DIR_REVERSE ) { // TODO - alarm w/ some persistence - SET_ALARM_WITH_2_U32_DATA( ALARM_ID_HD_SYRINGE_PUMP_CONTROLLER_DIRECTION_ERROR, (U32)syringePumpControllerMeasuredDirection, (U32)SYRINGE_PUMP_RETRACT_STATE ); +// SET_ALARM_WITH_2_U32_DATA( ALARM_ID_HD_SYRINGE_PUMP_CONTROLLER_DIRECTION_ERROR, (U32)syringePumpControllerMeasuredDirection, (U32)SYRINGE_PUMP_RETRACT_STATE ); } else if ( syringePumpEncoderMeasuredDirection != MOTOR_DIR_REVERSE ) { // TODO - alarm w/ some persistence - SET_ALARM_WITH_2_U32_DATA( ALARM_ID_HD_SYRINGE_PUMP_ENCODER_DIRECTION_ERROR, (U32)syringePumpEncoderMeasuredDirection, (U32)SYRINGE_PUMP_RETRACT_STATE ); +// SET_ALARM_WITH_2_U32_DATA( ALARM_ID_HD_SYRINGE_PUMP_ENCODER_DIRECTION_ERROR, (U32)syringePumpEncoderMeasuredDirection, (U32)SYRINGE_PUMP_RETRACT_STATE ); } else if ( ( TRUE == syringePumpPositionKnown ) && ( getSyringePumpPosition() < SYRINGE_PUMP_RETRACT_POS_MIN ) ) {