Index: firmware/App/Controllers/SyringePump.c =================================================================== diff -u -r825a299f4ad0f4a51b203c3f5d730c6978ab6844 -r6f1e959a3c5f991d7b8a787939d252c7998cbbd1 --- firmware/App/Controllers/SyringePump.c (.../SyringePump.c) (revision 825a299f4ad0f4a51b203c3f5d730c6978ab6844) +++ firmware/App/Controllers/SyringePump.c (.../SyringePump.c) (revision 6f1e959a3c5f991d7b8a787939d252c7998cbbd1) @@ -69,7 +69,7 @@ #define SYRINGE_PUMP_RATE_ALARM_PERSISTENCE ( MS_PER_SECOND / TASK_PRIORITY_INTERVAL ) ///< Alarm persistence period for syringe pump speed check alarms. #define SYRINGE_PUMP_DIR_ALARM_PERSISTENCE ( MS_PER_SECOND / TASK_PRIORITY_INTERVAL ) ///< Alarm persistence period for syringe pump direction check alarms. -#define SYRINGE_PUMP_OFF_ALARM_PERSISTENCE ( 500 / TASK_PRIORITY_INTERVAL ) ///< Alarm persistence period for syringe pump off check alarms. +#define SYRINGE_PUMP_OFF_ALARM_PERSISTENCE ( 1100 / TASK_PRIORITY_INTERVAL ) ///< Alarm persistence period for syringe pump off check alarms. #define SYRINGE_PUMP_OCCLUSION_ALARM_PERSISTENCE ( 30 / TASK_PRIORITY_INTERVAL ) ///< Alarm persistence period for syringe pump occlusion alarms. #define SYRINGE_PUMP_ADC_READ_PERSISTENCE ( 100 / TASK_PRIORITY_INTERVAL ) ///< Syringe pump ADC stale read alarm persistence time (100 ms).