Index: firmware/App/Controllers/SyringePump.c =================================================================== diff -u -rc8da2f1848da1dfb3ca1c4f2bb453a1f7d149b2e -rcc7f1bd82e346521f0102b3398bd030fb3bb0e6f --- firmware/App/Controllers/SyringePump.c (.../SyringePump.c) (revision c8da2f1848da1dfb3ca1c4f2bb453a1f7d149b2e) +++ firmware/App/Controllers/SyringePump.c (.../SyringePump.c) (revision cc7f1bd82e346521f0102b3398bd030fb3bb0e6f) @@ -68,7 +68,6 @@ #define SYRINGE_PUMP_RATE_ALARM_PERSISTENCE 3000 ///< Alarm persistence period (in ms) for syringe pump speed check alarms. #define SYRINGE_PUMP_DIR_ALARM_PERSISTENCE 3000 ///< Alarm persistence period (in ms) for syringe pump direction check alarms. #define SYRINGE_PUMP_OFF_ALARM_PERSISTENCE 1000 ///< Alarm persistence period (in ms) for syringe pump off check alarms. -#define SYRINGE_PUMP_OCCLUSION_ALARM_PERSISTENCE 30 ///< Alarm persistence period (in ms) for syringe pump occlusion alarms. #define SYRINGE_PUMP_ADC_READ_PERSISTENCE 100 ///< Syringe pump ADC stale read alarm persistence time (in ms). #define STEPS_TO_MICROSTEPS( s ) ( (s) * 32.0F ) ///< Macro conversion from steps to microsteps. @@ -91,6 +90,7 @@ /// Expected position of empty in relation to home position. #define SYRINGE_PUMP_EMPTY_POS ( SYRINGE_ENCODER_COUNTS_PER_ML * 10.84F ) + /// Margin of error for empty position determination. #define SYRINGE_PUMP_EMPTY_POS_MARGIN ( SYRINGE_ENCODER_COUNTS_PER_ML * 0.5F ) /// Minimum retract position. @@ -317,7 +317,6 @@ initPersistentAlarm( ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR, 0, SYRINGE_PUMP_OFF_ALARM_PERSISTENCE ); initPersistentAlarm( ALARM_ID_HD_SYRINGE_PUMP_NOT_STOPPED_ERROR, 0, SYRINGE_PUMP_OFF_ALARM_PERSISTENCE ); initPersistentAlarm( ALARM_ID_HD_SYRINGE_PUMP_SPEED_ERROR, 0, SYRINGE_PUMP_RATE_ALARM_PERSISTENCE ); - initPersistentAlarm( ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION, 0, SYRINGE_PUMP_OCCLUSION_ALARM_PERSISTENCE ); // Reset request flags resetSyringePumpRequestFlags(); @@ -1573,10 +1572,9 @@ F32 forceDelta = forceAtEndOfPriming - forceAtEndOfSeek; // Occlusion is detected if force at end of prime is > than force at end of seek by 0.5 volts or more BOOL occlusionDetected = ( forceDelta >= SYRINGE_FORCE_OCCLUSION_DIFF_V ? TRUE : FALSE ); - if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION, occlusionDetected ) ) + if ( TRUE == occlusionDetected ) { SET_ALARM_WITH_2_F32_DATA( ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION, forceAtEndOfSeek, forceAtEndOfPriming ) // reuse this alarm after prime - forceAtEndOfSeek = 0.0; // reset this value in case we want to prime again result = TRUE; } @@ -1885,6 +1883,7 @@ *************************************************************************/ + /*********************************************************************//** * @brief * The testSetSyringePumpDataPublishIntervalOverride function overrides the