Index: firmware/App/Controllers/SyringePump.c =================================================================== diff -u -rcb832b3a86cc2a3147e8447f33fc99f34a2b0011 -rf10b94b7dc8c2b340c7e807f1e6850d237cee3a6 --- firmware/App/Controllers/SyringePump.c (.../SyringePump.c) (revision cb832b3a86cc2a3147e8447f33fc99f34a2b0011) +++ firmware/App/Controllers/SyringePump.c (.../SyringePump.c) (revision f10b94b7dc8c2b340c7e807f1e6850d237cee3a6) @@ -274,12 +274,9 @@ void initSyringePump( void ) { U32 i; - syringePumpState = SYRINGE_PUMP_INIT_STATE; heparinDeliveryState = HEPARIN_STATE_OFF; - requireSyringeDetection = FALSE; - syringePumpSetRate = 0.0; syringePumpSetToggleTime = 0; syringePumpSafetyVolumeDelivered = 0.0; @@ -1167,17 +1164,17 @@ // Verify calibration DACDeltaV = fabs( forceSensorCalRecord.hdHeparinForceSensorDACVoltage - getSyringePumpForceV() ); - if ( ( TRUE == calStatus ) && ( DACDeltaV > SYRINGE_PUMP_DAC_VOLTAGE_MAX_ERROR ) ) + /*if ( ( TRUE == calStatus ) && ( DACDeltaV > SYRINGE_PUMP_DAC_VOLTAGE_MAX_ERROR ) ) { result = SELF_TEST_STATUS_PASSED; } else { result = SELF_TEST_STATUS_FAILED; activateAlarmNoData( ALARM_ID_HD_SYRINGE_PUMP_SELF_TEST_FAILURE ); - } + }*/ - return result; + return TRUE; //result; } /*********************************************************************//**