Index: firmware/App/Controllers/SyringePump.c =================================================================== diff -u -rdb1f997cc2b3abe3a3daf91e93e5bd6d3f8fb5ed -r88edec47284a01d54fa711109ff3ba1277ab1a2e --- firmware/App/Controllers/SyringePump.c (.../SyringePump.c) (revision db1f997cc2b3abe3a3daf91e93e5bd6d3f8fb5ed) +++ firmware/App/Controllers/SyringePump.c (.../SyringePump.c) (revision 88edec47284a01d54fa711109ff3ba1277ab1a2e) @@ -1623,7 +1623,7 @@ if ( TRUE == didTimeout( syringePumpStateStartTime, SYRINGE_PUMP_PRIMING_TIMEOUT_MS ) ) { stopPump = TRUE; - SET_ALARM_WITH_2_F32_DATA( ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION, (F32)SYRINGE_PUMP_PRIMING_TIMEOUT_MS, 0.0F ) + SET_ALARM_WITH_2_F32_DATA( ALARM_ID_HD_SYRINGE_PUMP_PRIME_TIMEOUT, (F32)SYRINGE_PUMP_PRIMING_TIMEOUT_MS, 0.0F ) } // Check for stall