Index: firmware/App/Controllers/SyringePump.c =================================================================== diff -u -rf10b94b7dc8c2b340c7e807f1e6850d237cee3a6 -r3c5b521fcaa489eef9df0e2d657c7e0b23dd8ece --- firmware/App/Controllers/SyringePump.c (.../SyringePump.c) (revision f10b94b7dc8c2b340c7e807f1e6850d237cee3a6) +++ firmware/App/Controllers/SyringePump.c (.../SyringePump.c) (revision 3c5b521fcaa489eef9df0e2d657c7e0b23dd8ece) @@ -1164,17 +1164,17 @@ // Verify calibration DACDeltaV = fabs( forceSensorCalRecord.hdHeparinForceSensorDACVoltage - getSyringePumpForceV() ); - /*if ( ( TRUE == calStatus ) && ( DACDeltaV > SYRINGE_PUMP_DAC_VOLTAGE_MAX_ERROR ) ) + if ( ( TRUE == calStatus ) && ( DACDeltaV > SYRINGE_PUMP_DAC_VOLTAGE_MAX_ERROR ) ) { result = SELF_TEST_STATUS_PASSED; } else { result = SELF_TEST_STATUS_FAILED; activateAlarmNoData( ALARM_ID_HD_SYRINGE_PUMP_SELF_TEST_FAILURE ); - }*/ + } - return TRUE; //result; + return result; } /*********************************************************************//**