Index: firmware/App/HDCommon.h =================================================================== diff -u -r372e1d2b228b6c29a17b79b241d2f6f94bf1e801 -rfb60e77d7cf6cda1385edc0d07a44790b9148a67 --- firmware/App/HDCommon.h (.../HDCommon.h) (revision 372e1d2b228b6c29a17b79b241d2f6f94bf1e801) +++ firmware/App/HDCommon.h (.../HDCommon.h) (revision fb60e77d7cf6cda1385edc0d07a44790b9148a67) @@ -50,10 +50,10 @@ // #define RUN_DPI_OPEN_LOOP 1 // Run dialysate inlet pump in open loop mode #define WORN_OUT_CARTRIDGE 1 // Running with an old worn out cartridge (max wear) #define PBA_ESTIMATION 1 // Estimate arterial pressure rather than look at PBA sensor -// #define DISABLE_MOTOR_CURRENT_CHECKS 1 // Do not error on HD pump current checks -// #define DISABLE_PUMP_FLOW_CHECKS 1 // Do not error on HD pump flow checks -// #define DISABLE_PUMP_SPEED_CHECKS 1 // Do not error on HD pump speed checks -// #define DISABLE_PUMP_DIRECTION_CHECKS 1 // Do not error on HD pump direction checks + #define DISABLE_MOTOR_CURRENT_CHECKS 1 // Do not error on HD pump current checks + #define DISABLE_PUMP_FLOW_CHECKS 1 // Do not error on HD pump flow checks + #define DISABLE_PUMP_SPEED_CHECKS 1 // Do not error on HD pump speed checks + #define DISABLE_PUMP_DIRECTION_CHECKS 1 // Do not error on HD pump direction checks // #define DISABLE_SYRINGE_PUMP 1 // Disable syringe pump functionality #define ALWAYS_ALLOW_SYRINGE_PUMP_CMDS 1 // Allow syringe pump commands at any time except when pump is busy #define DISABLE_PRESSURE_CHECKS 1 // Do not error on HD pressure checks