Index: firmware/App/HDCommon.h =================================================================== diff -u -r946f42bfd6606dd3d771d9a4a7b50e5678469f62 -r7a1358765fb2b530e505ac3ce7ceecfb45f8c6ee --- firmware/App/HDCommon.h (.../HDCommon.h) (revision 946f42bfd6606dd3d771d9a4a7b50e5678469f62) +++ firmware/App/HDCommon.h (.../HDCommon.h) (revision 7a1358765fb2b530e505ac3ce7ceecfb45f8c6ee) @@ -58,7 +58,7 @@ #define ALWAYS_ALLOW_SYRINGE_PUMP_CMDS 1 // Allow syringe pump commands at any time except when pump is busy #define DISABLE_PRESSURE_CHECKS 1 // Do not error on HD pressure checks #define DISABLE_ARTERIAL_PRESSURE_CHECK 1 // Do not error on arterial pressure out of range - #define DISABLE_VENOUS_PRESSURE_CHECK 1 // Do not error on venous pressure out of range +// #define DISABLE_VENOUS_PRESSURE_CHECK 1 // Do not error on venous pressure out of range // #define DISABLE_UF_ALARMS 1 // Do not error on HD ultrafiltration checks #define DISABLE_DIALYSATE_TEMP_CHECK 1 // Disable dialysate temperature check // #define DISABLE_VALVE_ALARMS 1 // Do not error on HD valve position @@ -77,7 +77,7 @@ #define ALLOW_1_MIN_TREATMENT_DURATION 1 // Allow user to change treatment duration to as low as 1 minute #define DISABLE_SYRINGE_PUMP_ALARMS 1 // Disable some syringe pump alarms that are triggering intermittently // #define NO_PUMP_FLOW_LIMITS 1 // Allow any commanded flow rate for peristaltic pumps - #define DISABLE_BUBBLE_ALARMS 1 // Disable bubble alarms +// #define DISABLE_BUBBLE_ALARMS 1 // Disable bubble alarms #define DISABLE_UI_POST_TEST 1 // Disable the UI POST // #define DISABLE_UI_COMM_TO_ALARM 1 // Disable UI comm timeouts #define DISABLE_WD_AND_SFTY_POST_TESTS 1 // Disable watchdog and safety shutdown POST tests