Index: firmware/App/HDCommon.h =================================================================== diff -u -ra3e8170ae9c98e957fdc8cd17c3ba2c3f3baec01 -r708c23e585468515a8467d0056d712b2846b56da --- firmware/App/HDCommon.h (.../HDCommon.h) (revision a3e8170ae9c98e957fdc8cd17c3ba2c3f3baec01) +++ firmware/App/HDCommon.h (.../HDCommon.h) (revision 708c23e585468515a8467d0056d712b2846b56da) @@ -66,8 +66,8 @@ // #define TEMP_UI_ALARM_SILENCE_FIX 1 // Temporary UI fix for handling alarm silence request msg // #define SKIP_PRIMING 1 // Skip Pre-treatment Prime #define SKIP_WET_SELF_TESTS 1 // Skip Pre-treatment prime wet self-tests -// #define V2_0_SYSTEM 1 // Build for v2.0 system -// #define SKIP_SAMPLE_WATER 1 // Skip pre-treatment sample water + #define V2_0_SYSTEM 1 // Build for v2.0 system + #define SKIP_SAMPLE_WATER 1 // Skip pre-treatment sample water #define SKIP_CONSUMABLE_TESTS 1 // Skip pre-treatment consumable Self-tests #define SKIP_DRY_SELF_TESTS 1 // Skip pre-treatment dry self-tests // #define SKIP_UI_INTERACTION 1 // Skip UI interaction. @@ -79,7 +79,8 @@ // #define DISABLE_FPGA_COUNTER_CHECKS 1 // Disable alarms associated with FPGA read/error counters // #define DISABLE_VOLTAGE_MONITOR 1 // Disable voltage monitoring/alarms #define ALLOW_1_MIN_TREATMENT_DURATION 1 // Allow user to change treatment duration to as low as 1 minute -// #define DISABLE_SYRINGE_PUMP_ALARMS 1 // Disable some syringe pump alarms that are triggering intermittently + #define DISABLE_SYRINGE_PUMP_ALARMS 1 // Disable some syringe pump alarms that are triggering intermittently + #define NO_PUMP_FLOW_LIMITS 1 // Allow any commanded flow rate for peristaltic pumps #include #include