Index: firmware/App/HDCommon.h =================================================================== diff -u -rc239550921da42b0c73d4ff761022f44a990fb3b -rba60637ad5158bd50803aa55408242e68429e6b1 --- firmware/App/HDCommon.h (.../HDCommon.h) (revision c239550921da42b0c73d4ff761022f44a990fb3b) +++ firmware/App/HDCommon.h (.../HDCommon.h) (revision ba60637ad5158bd50803aa55408242e68429e6b1) @@ -50,10 +50,10 @@ // #define RUN_DPI_OPEN_LOOP 1 // Run dialysate inlet pump in open loop mode #define WORN_OUT_CARTRIDGE 1 // Running with an old worn out cartridge (max wear) #define PBA_ESTIMATION 1 // Estimate arterial pressure rather than look at PBA sensor -// #define DISABLE_MOTOR_CURRENT_CHECKS 1 // Do not error on HD pump current checks -// #define DISABLE_PUMP_FLOW_CHECKS 1 // Do not error on HD pump flow checks -// #define DISABLE_PUMP_SPEED_CHECKS 1 // Do not error on HD pump speed checks -// #define DISABLE_PUMP_DIRECTION_CHECKS 1 // Do not error on HD pump direction checks + #define DISABLE_MOTOR_CURRENT_CHECKS 1 // Do not error on HD pump current checks + #define DISABLE_PUMP_FLOW_CHECKS 1 // Do not error on HD pump flow checks + #define DISABLE_PUMP_SPEED_CHECKS 1 // Do not error on HD pump speed checks + #define DISABLE_PUMP_DIRECTION_CHECKS 1 // Do not error on HD pump direction checks // #define DISABLE_SYRINGE_PUMP 1 // Disable syringe pump functionality #define ALWAYS_ALLOW_SYRINGE_PUMP_CMDS 1 // Allow syringe pump commands at any time except when pump is busy #define DISABLE_PRESSURE_CHECKS 1 // Do not error on HD pressure checks