Index: firmware/App/HDCommon.h =================================================================== diff -u -rc39d7514c94fd61972b6e05e91166fea5b645373 -rc2ca8fef01dfd51d6b7bfe47bc7397a530428d09 --- firmware/App/HDCommon.h (.../HDCommon.h) (revision c39d7514c94fd61972b6e05e91166fea5b645373) +++ firmware/App/HDCommon.h (.../HDCommon.h) (revision c2ca8fef01dfd51d6b7bfe47bc7397a530428d09) @@ -77,7 +77,7 @@ #define ALLOW_1_MIN_TREATMENT_DURATION 1 // Allow user to change treatment duration to as low as 1 minute #define DISABLE_SYRINGE_PUMP_ALARMS 1 // Disable some syringe pump alarms that are triggering intermittently // #define NO_PUMP_FLOW_LIMITS 1 // Allow any commanded flow rate for peristaltic pumps -// #define DISABLE_BUBBLE_ALARMS 1 // Disable bubble alarms + #define DISABLE_BUBBLE_ALARMS 1 // Disable bubble alarms #define DISABLE_UI_POST_TEST 1 // Disable the UI POST // #define DISABLE_UI_COMM_TO_ALARM 1 // Disable UI comm timeouts #define DISABLE_WD_AND_SFTY_POST_TESTS 1 // Disable watchdog and safety shutdown POST tests