Index: firmware/App/Modes/BloodPrime.c =================================================================== diff -u -r9d924ec76ce770eb2a94088bfe37bfb0e30981a8 -r41eaf78372c0ee2a79b38d0a06bd3f02d54baed4 --- firmware/App/Modes/BloodPrime.c (.../BloodPrime.c) (revision 9d924ec76ce770eb2a94088bfe37bfb0e30981a8) +++ firmware/App/Modes/BloodPrime.c (.../BloodPrime.c) (revision 41eaf78372c0ee2a79b38d0a06bd3f02d54baed4) @@ -37,7 +37,6 @@ // ********** private definitiions *********** #define TUBING_BLOOD_PRIME_VOLUME_ML 80.0F ///< Target blood prime volume to prime the blood side circuit (in mL). -#define MAX_BLOOD_PRIME_VOLUME_ERROR_ML 60.0F ///< Maximum error in total additional blood prime volume (20% of total). #define MIN_RAMP_TIME_SEC 60 ///< Minimum ramp time for blood prime (in seconds). /// Initial flow rate for blood pump when starting blood prime operation. @@ -66,7 +65,6 @@ static OVERRIDE_U32_T bloodPrimePublishInterval = { BLOOD_PRIME_DATA_PUBLISH_INTERVAL, BLOOD_PRIME_DATA_PUBLISH_INTERVAL, BLOOD_PRIME_DATA_PUBLISH_INTERVAL, 0 }; static OVERRIDE_F32_T cumulativeBloodPrimeVolume_mL = { 0.0, 0.0, 0.0, 0 }; ///< Total cumulative blood prime volume (in mL) from measured blood flow rate. -static F32 expectedBloodPrimeVolume_mL = 0.0; ///< Total cumulative blood prime volume (in mL) expected based on target blood flow rate. // ********** private function prototypes ********** @@ -92,7 +90,6 @@ bloodPrimeRampControlTimerCtr = 0; bloodPrimePublishTimerCtr = BLOOD_PRIME_DATA_PUBLISH_INTERVAL - 2; // setup so publish will occur time after next cumulativeBloodPrimeVolume_mL.data = 0.0; - expectedBloodPrimeVolume_mL = 0.0; resetBloodPrimeFlags(); bloodPrimeTargetVolume_mL = (F32)TUBING_BLOOD_PRIME_VOLUME_ML + (F32)getDialyzerBloodVolume(); @@ -226,7 +223,6 @@ // Update blood prime volume delivered so far cumulativeBloodPrimeVolume_mL.data += ( getMeasuredBloodFlowRate() * BLOOD_PRIME_FLOW_INTEGRATOR ); - expectedBloodPrimeVolume_mL += ( (F32)getTargetBloodFlowRate() * BLOOD_PRIME_FLOW_INTEGRATOR ); // Has blood prime completed? if ( getBloodPrimeVolume() >= bloodPrimeTargetVolume_mL ) @@ -237,16 +233,6 @@ signalBloodPrimeToDialysis(); // Signal treatment mode that it's time to start dialysis cmdStopDGTrimmerHeater(); // Stop trimmer heater - dialysis sub-mode will restart as appropriate } - // Is blood prime taking too long based on set BP rate? - else if ( fabs( expectedBloodPrimeVolume_mL - getBloodPrimeVolume() ) > MAX_BLOOD_PRIME_VOLUME_ERROR_ML ) - { -#ifndef _RELEASE_ - if ( getSoftwareConfigStatus( SW_CONFIG_DISABLE_PUMP_FLOW_CHECKS ) != SW_CONFIG_ENABLE_VALUE ) -#endif - { - activateAlarmNoData( ALARM_ID_BLOOD_PRIME_VOLUME_CHECK_FAILURE ); - } - } else { // ramp blood pump on ramp interval