Index: firmware/App/Modes/BloodPrime.c =================================================================== diff -u -rc65ad0538ff99c3e13d7d7866ac15e38a1ef6002 -r8466e63f95f65a3ffb18c3af85ac99328e41167b --- firmware/App/Modes/BloodPrime.c (.../BloodPrime.c) (revision c65ad0538ff99c3e13d7d7866ac15e38a1ef6002) +++ firmware/App/Modes/BloodPrime.c (.../BloodPrime.c) (revision 8466e63f95f65a3ffb18c3af85ac99328e41167b) @@ -39,10 +39,13 @@ // TODO - get from Systems when available #define TARGET_BLOOD_PRIME_VOLUME_ML 300.0 ///< Target blood prime volume to prime the blood side circuit (in mL). #define MIN_RAMP_TIME_SEC 60 ///< Minimum ramp time for blood prime (in seconds). + +#ifndef DISABLE_PUMP_FLOW_CHECKS /// Maximum blood prime volume measured by independent means (as % of target). static const F32 MAX_BLOOD_PRIME_SAFETY_VOLUME_ML = ( TARGET_BLOOD_PRIME_VOLUME_ML * 1.2 ); /// Minimum blood prime volume measured by independent means (as % of target). static const F32 MIN_BLOOD_PRIME_SAFETY_VOLUME_ML = ( TARGET_BLOOD_PRIME_VOLUME_ML * 0.8 ); +#endif /// Initial flow rate for blood pump when starting blood prime operation. #define BLOOD_PRIME_INIT_BP_FLOW_RATE_ML_MIN 100