Index: firmware/App/Modes/BloodPrime.c =================================================================== diff -u -rd33ae635eb38ae8f2647f68304be74e8e4f78cfd -r667feab319b14681a0330a715da7a1ba930b518d --- firmware/App/Modes/BloodPrime.c (.../BloodPrime.c) (revision d33ae635eb38ae8f2647f68304be74e8e4f78cfd) +++ firmware/App/Modes/BloodPrime.c (.../BloodPrime.c) (revision 667feab319b14681a0330a715da7a1ba930b518d) @@ -235,14 +235,12 @@ signalBloodPrimeToDialysis(); // Signal treatment mode that it's time to start dialysis cmdStopDGTrimmerHeater(); // Stop trimmer heater - dialysis sub-mode will restart as appropriate } -#ifndef DISABLE_PUMP_FLOW_CHECKS // Is blood prime taking too long based on set BP rate? else if ( ( fabs( expectedBloodPrimeVolume_mL - getBloodPrimeVolume() ) > MAX_BLOOD_PRIME_VOLUME_ERROR_ML ) && ( getSoftwareConfigStatus( SW_CONFIG_DISABLE_PUMP_FLOW_CHECKS ) != SW_CONFIG_ENABLE_VALUE ) ) { activateAlarmNoData( ALARM_ID_BLOOD_PRIME_VOLUME_CHECK_FAILURE ); } -#endif else { // ramp blood pump on ramp interval