Index: firmware/App/Modes/Dialysis.c =================================================================== diff -u -r37a9fd8f15e413db5337371a7d1a1cb65567af7c -r2f4c3f2e85936ac4dcf3e25d1e816b38a970c484 --- firmware/App/Modes/Dialysis.c (.../Dialysis.c) (revision 37a9fd8f15e413db5337371a7d1a1cb65567af7c) +++ firmware/App/Modes/Dialysis.c (.../Dialysis.c) (revision 2f4c3f2e85936ac4dcf3e25d1e816b38a970c484) @@ -515,7 +515,7 @@ setDialOutPumpTargetRate( setDialysateFlowRate, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_CLOSED_LOOP ); // Go to UF paused state currentUFState = UF_PAUSED_STATE; - sendTreatmentLogEventData( UF_PAUSE_RESUME_EVENT, UF_RUNNING_STATE, UF_PAUSED_STATE ); + sendTreatmentLogEventData( UF_START_RESUME_EVENT, UF_RUNNING_STATE, UF_PAUSED_STATE ); } else { @@ -567,7 +567,7 @@ lastUFTimeStamp = getMSTimerCount(); // Go to UF running state currentUFState = UF_RUNNING_STATE; - sendTreatmentLogEventData( UF_PAUSE_RESUME_EVENT, UF_PAUSED_STATE, UF_RUNNING_STATE ); + sendTreatmentLogEventData( UF_START_RESUME_EVENT, UF_PAUSED_STATE, UF_RUNNING_STATE ); } else {