Index: firmware/App/Modes/Dialysis.c =================================================================== diff -u -r79cd16f091997683de89e2e84d3a86e2dd7d5493 -r8b824fc931d5af91a791644be64f6ad5b1e7ae7f --- firmware/App/Modes/Dialysis.c (.../Dialysis.c) (revision 79cd16f091997683de89e2e84d3a86e2dd7d5493) +++ firmware/App/Modes/Dialysis.c (.../Dialysis.c) (revision 8b824fc931d5af91a791644be64f6ad5b1e7ae7f) @@ -40,8 +40,8 @@ // ********** private definitions ********** -#define MAX_UF_VOLUME_ACCURACY_ERROR_ML 100.0F ///< Maximum ultrafiltration volume accuracy error in mL over the entire treatment. -#define MAX_UF_RATE_ACCURACY_ERROR_ML_MIN 10.0F ///< Maximum ultrafiltration rate accuracy error in mL/min over use of single reservoir during treatment. +#define MAX_UF_VOLUME_ACCURACY_ERROR_ML 100.0F ///< Maximum ultrafiltration volume accuracy error in mL over the entire treatment. +#define LOAD_CELL_PRIMARY_BACKUP_MAX_ALLOWED_DRIFT_GRAMS 20.0F ///< Reservoir load cell drift difference allowed /// Saline bolus data broadcast interval (ms/task time) count. static const U32 SALINE_BOLUS_DATA_PUB_INTERVAL = ( MS_PER_SECOND / TASK_GENERAL_INTERVAL );