Index: firmware/App/Modes/Dialysis.c =================================================================== diff -u -rce3e0696642099164fa482c864509c67ce65579b -r6a414654391c1729525d55812abbf111ddef7d6a --- firmware/App/Modes/Dialysis.c (.../Dialysis.c) (revision ce3e0696642099164fa482c864509c67ce65579b) +++ firmware/App/Modes/Dialysis.c (.../Dialysis.c) (revision 6a414654391c1729525d55812abbf111ddef7d6a) @@ -1062,16 +1062,8 @@ *************************************************************************/ static void checkUFAccuracyAndVolume( void ) { - F32 uFMeasRate = measUFVolume - lastUFVolumeChecked; + F32 uFMeasRate = measUFVolume - lastUFVolumeChecked; // volumes are at start/end of 1 hour period, so implied rate is per hour - // check max UF rate in last hour // TODO - is this check necessary considering next check below is a tighter check on this? -// if ( uFMeasRate > MAX_UF_RATE_ML_PER_HOUR ) -// { -//#ifndef DISABLE_UF_ALARMS -// SET_ALARM_WITH_1_F32_DATA( ALARM_ID_UF_RATE_TOO_HIGH_ERROR, uFMeasRate ); -//#endif -// } - // check UF rate diff from target in last hour if ( uFMeasRate > maxUFRateAccuracyError_Ml_hr ) {