Index: firmware/App/Modes/ModeInitPOST.c =================================================================== diff -u -r19a8bf98a7154e24c35da25225d4b55bf70ddd09 -rcbaa57803fa8359407a799fca0cde6ae2c486bd7 --- firmware/App/Modes/ModeInitPOST.c (.../ModeInitPOST.c) (revision 19a8bf98a7154e24c35da25225d4b55bf70ddd09) +++ firmware/App/Modes/ModeInitPOST.c (.../ModeInitPOST.c) (revision cbaa57803fa8359407a799fca0cde6ae2c486bd7) @@ -8,7 +8,7 @@ * @file ModeInitPOST.c * * @author (last) Dara Navaei -* @date (last) 03-Apr-2022 +* @date (last) 13-Jul-2022 * * @author (original) Dara Navaei * @date (original) 05-Nov-2019 @@ -156,14 +156,6 @@ postState = handlePOSTStatus( testStatus ); break; - case POST_STATE_AC: - if ( getMonitoredLineLevel( MONITORED_LINE_24V ) < MIN_24V_MEASURED_FOR_AC ) - { - initiatePowerOff(); - } - postState = POST_STATE_WATCHDOG; - break; - case POST_STATE_WATCHDOG: testStatus = execWatchdogTest(); postState = handlePOSTStatus( testStatus ); @@ -191,6 +183,7 @@ case POST_STATE_BLOOD_FLOW: testStatus = execBloodFlowTest(); postState = handlePOSTStatus( testStatus ); + sendUIRequestPOSTFinalResult(); // request UI POST final result here so we have it before UI test below break; case POST_STATE_DIALYSATE_INLET_FLOW: