Index: firmware/App/Modes/ModePostTreat.c =================================================================== diff -u -ra74a984a7059f75d86ad87d6d9499bd8f94cc976 -r57ee0134869672b53ab5b7146b8988ede8f828d6 --- firmware/App/Modes/ModePostTreat.c (.../ModePostTreat.c) (revision a74a984a7059f75d86ad87d6d9499bd8f94cc976) +++ firmware/App/Modes/ModePostTreat.c (.../ModePostTreat.c) (revision 57ee0134869672b53ab5b7146b8988ede8f828d6) @@ -56,10 +56,13 @@ *************************************************************************/ void transitionToPostTreatmentMode( void ) { - // temporary test code - alarm lamp high alarm + // temporary test code. TODO - remove later +#ifndef UF_TEST_ENABLED setBloodPumpTargetFlowRate( 100, MOTOR_DIR_REVERSE, PUMP_CONTROL_MODE_OPEN_LOOP ); setDialInPumpTargetFlowRate( 100, MOTOR_DIR_REVERSE, PUMP_CONTROL_MODE_OPEN_LOOP ); setDialOutPumpTargetRate( 100, MOTOR_DIR_REVERSE, PUMP_CONTROL_MODE_OPEN_LOOP ); +#else +#endif #ifdef RM46_EVAL_BOARD_TARGET start = getMSTimerCount(); #endif @@ -78,7 +81,9 @@ { BOOL stop = isStopButtonPressed(); +#ifndef UF_TEST_ENABLED if ( TRUE == stop ) +#endif { requestNewOperationMode( MODE_STAN ); }