Index: firmware/App/Modes/ModePreTreat.c =================================================================== diff -u -r06eb4485aab0b6f160610358395d8a568f239ed1 -re408267b7e73d664051a46e65ac248749af8e5be --- firmware/App/Modes/ModePreTreat.c (.../ModePreTreat.c) (revision 06eb4485aab0b6f160610358395d8a568f239ed1) +++ firmware/App/Modes/ModePreTreat.c (.../ModePreTreat.c) (revision e408267b7e73d664051a46e65ac248749af8e5be) @@ -629,8 +629,7 @@ signalBloodPumpHardStop(); signalDialOutPumpHardStop(); - //setDialInPumpTargetFlowRate( DIP_PATIENT_CONNECTION_FLOW_RATE_ML_MIN, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); - setDialInPumpTargetFlowRate( 250, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); // TODO remove this line once the new flow control is implemented + setDialInPumpTargetFlowRate( DIP_PATIENT_CONNECTION_FLOW_RATE_ML_MIN, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); cmdStartDGTrimmerHeater(); } @@ -1175,6 +1174,7 @@ rsrvrCmd.reservoirID = (U32)DG_RESERVOIR_1; rsrvrCmd.useLastTrimmerHeaterDC = FALSE; reservoirFlushedStatus[ DG_RESERVOIR_1 ] = TRUE; + cmdSetDGActiveReservoir( &rsrvrCmd ); } } @@ -1185,6 +1185,7 @@ rsrvrCmd.reservoirID = (U32)DG_RESERVOIR_2; rsrvrCmd.useLastTrimmerHeaterDC = FALSE; reservoirFlushedStatus[ DG_RESERVOIR_2 ] = TRUE; + cmdSetDGActiveReservoir( &rsrvrCmd ); } }